Evaluation of the impeller shroud performance of an axial flow ventricular assist device using computational fluid dynamics.

Boyang Su1, Leok P Chua, Tau M Lim, Tongming Zhou.   

Abstract

Generally, there are two types of impeller design used in the axial flow blood pumps. For the first type, which can be found in most of the axial flow blood pumps, the magnet is embedded inside the impeller hub or blades. For the second type, the magnet is embedded inside the cylindrical impeller shroud, and this design has not only increased the rotating stability of the impeller but has also avoided the flow interaction between the impeller blade tip and the pump casing. Although the axial flow blood pumps with either impeller design have been studied individually, the comparisons between these two designs have not been conducted in the literature. Therefore, in this study, two axial flow blood pumps with and without impeller shrouds were numerically simulated with computational fluid dynamics and compared with each other in terms of hydraulic and hematologic performances. For the ease of comparison, these two models have the same inner components, which include a three-blade straightener, a two-blade impeller, and a three-blade diffuser. The simulation results showed that the model with impeller shroud had a lower static pressure head with a lower hydraulic efficiency than its counterpart. It was also found that the blood had a high possibility to deposit on the impeller shroud inner surface, which greatly enhanced the possibility of thrombus formation. The blood damage indices in both models were around 1%, which was much lower than the 13.1% of the axial flow blood pump of Yano et al. with the corresponding experimental hemolysis of 0.033 g/100 L.
© 2010, Copyright the Authors. Artificial Organs © 2010, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20883393     DOI: 10.1111/j.1525-1594.2010.01099.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  Development of a novel shrouded impeller pediatric blood pump.

Authors:  Talha Irfan Khan; Haris Sheh Zad; Ismail Lazoglu; Ozlem Yalcin
Journal:  J Artif Organs       Date:  2018-02-24       Impact factor: 1.731

2.  Hemolytic Performance in Two Generations of the Sputnik Left Ventricular Assist Device: A Combined Numerical and Experimental Study.

Authors:  Alexandra N Romanova; Alexander A Pugovkin; Maxim V Denisov; Ivan A Ephimov; Dmitry V Gusev; Marian Walter; Thomas Groth; Olga L Bockeria; Tatyana G Le; Anna S Satyukova; Sergey V Selishchev; Dmitry V Telyshev
Journal:  J Funct Biomater       Date:  2022-01-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.