Fully autonomous preload-sensitive control of implantable rotary blood pumps.

Andreas Arndt1, Peter Nüsser, Bernhard Lampe.   

Abstract

A pulsatility-based control algorithm with a self-adapting pulsatility reference value is proposed for an implantable rotary blood pump and is to be tested in computer simulations. The only input signal is the pressure difference across the pump, which is deduced from measurements of the pump's magnetic bearing. A pulsatility index (PI) is calculated as the mean absolute deviation from the mean pressure difference. As a second characteristic, the gradient of the PI with respect to the pump speed is derived. This pulsatility gradient (GPI) is used as the controlled variable to adjust the operating point of the pump when physiological variables such as the systemic arterial pressure, left ventricular contractility, or heart rate change. Depending on the selected mode of operation, the controller is either a linear controller or an extremum-seeking controller. A supervisory mechanism monitors the state of the system and projects the system into the region of convergence when necessary. The controller of the GPI continuously adjusts the reference value for PI. An underlying robust linear controller regulates the PI to the reference value in order to take into account changes in pulmonary venous return. As a means of reacting to sudden changes in the venous return, a suction detection mechanism was included. The control system is robustly stable within a wide range of physiological variables. All the clinician needs to do is to select between the two operating modes. No other adjustments are required. The algorithm showed promising results which encourage further testing in vitro and in vivo.
© 2010, Copyright the Authors. Artificial Organs © 2010, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20883392     DOI: 10.1111/j.1525-1594.2010.01092.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  4 in total

1.  Preload-based starling-like control for rotary blood pumps: numerical comparison with pulsatility control and constant speed operation.

Authors:  Mahdi Mansouri; Robert F Salamonsen; Einly Lim; Rini Akmeliawati; Nigel H Lovell
Journal:  PLoS One       Date:  2015-04-07       Impact factor: 3.240

2.  An Intra-Cycle Optimal Control Framework for Ventricular Assist Devices Based on Atrioventricular Plane Displacement Modeling.

Authors:  Clemens Zeile; Thomas Rauwolf; Alexander Schmeisser; Jeremi Kaj Mizerski; Rüdiger C Braun-Dullaeus; Sebastian Sager
Journal:  Ann Biomed Eng       Date:  2021-09-21       Impact factor: 3.934

3.  A Novel Control Method for Rotary Blood Pumps as Left Ventricular Assist Device Utilizing Aortic Valve State Detection.

Authors:  Dmitry Petukhov; Leonie Korn; Marian Walter; Dmitry Telyshev
Journal:  Biomed Res Int       Date:  2019-12-11       Impact factor: 3.411

4.  Improved Estimation of Left Ventricular Volume from Electric Field Modeling.

Authors:  Leonie Korn; Stephan Dahlmanns; Steffen Leonhardt; Marian Walter
Journal:  J Electr Bioimpedance       Date:  2021-12-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.