| Literature DB >> 20883388 |
Helen M Hayes1, Lawrence G Dembo, Robert Larbalestier, Gerry O'Driscoll.
Abstract
Gastrointestinal (GI) bleeding in ventricular assist devices (VADs) has been reported with rotary devices. The pathophysiological mechanisms and treatments are in evolution. We performed a retrospective review of GI bleeding episodes for all VADs implanted at our institution. Five male patients experienced GI bleeding-age 63.6 ± 3.64 years. VAD type VentrAssist n = 1, Jarvik 2000 n = 2, and HeartWare n = 2. All patients were anticoagulated as per protocol with antiplatelet agents (aspirin and/or clopidogrel bisulfate [Plavix] and warfarin (therapeutic international normalized ratio 2.0-3.5). There was no prior history of gastric bleeding in this group. Ten episodes of bleeding requiring blood transfusion occurred in five patients. Some patients had multiple episodes (1 × 5, 1 × 2, 3 × 1). The events occurred at varying times post-VAD implantation (days 14, 21, 26, 107, 152, 189, 476, 582, 669, and 839). Octreotide (a long-acting somatostatin analogue that reduces splanchnic arterial and portal blood flow) was administered subcutaneously or intravenously. Three patients received infusions of adrenaline at 1 µg/min to enhance pulsatility. Anticoagulation was interrupted during bleeding episodes but successfully introduced post bleeding event. GI bleeding is a significant complication of VAD therapy. In this article, we discuss diagnosis and management options.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20883388 DOI: 10.1111/j.1525-1594.2010.01084.x
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094