Literature DB >> 20879959

Maltose interference-free test strips for blood glucose testing at point-of-care: a laboratory performance evaluation.

Wai-Yoong Ng1, Chiau Chiae Tiong, Edward Jacob.   

Abstract

BACKGROUND: Maltose interference is a concern with blood glucose testing at point-of-care. We evaluated a maltose interference-free test strip (with a modified recombinant glucose dehydrogenase-pyrroloquinoline quinone system) for the Accu-Chek(®) Performa glucose meter (Roche Diagnostics, Mannheim, Germany).
METHODS: Blood specimens (n = 120) sent for clinical laboratory glucose testing were used in assessing performance characteristics, including imprecision, linearity, clinical impact analysis, and method comparison, of the test strips. To evaluate sugar interference, two heparinized blood specimens were spiked with maltose, xylose, and galactose (up to 500 mg/dL) followed by testing with modified Performa, Accutrend(®) (Roche Diagnostics), and Advantage II (Roche Diagnostics) test strips and by the laboratory method.
RESULTS: Test strips demonstrated total laboratory coefficients of variation of <7%; within-run coefficients of variation were 2.7-5.4% for blood glucose at 2.5-19.7 mmol/L. Clarke Error Grid analysis of the 120 results (0.8-27.6 mmol/L) showed all values to be within critical clinical limits. Comparison with laboratory results gave 0.960 correlation (Spearman's r(2)) with a Deming regression y (Performa) = 0.95x (laboratory) - 0.11 mmol/L (SEy|x0.06 mmol/L). A slight negative bias (-0.5 mmol/L) was demonstrated with the Bland-Altman difference plot. Maltose (up to 13.9 mmol/L) and xylose (33.3 mmol/L) had no effect, but galactose (2.2 mmol/L) showed interference. The sugars also affected test strips for Advantage II but not Accutrend glucose meters. With International Organization for Standardization ISO 15197:2003 criteria, 99% of the 120 results determined by the test strips were within the minimal acceptable performance; only one of 106 (5.9 mmol/L) was >20% from the laboratory result.
CONCLUSIONS: The modified and improved Performa test strips were not affected by maltose and xylose. They meet ISO 15197:2003 requirements with a slight bias (-0.5 mmol/L) compared to the laboratory method.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879959     DOI: 10.1089/dia.2010.0095

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  5 in total

Review 1.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

3.  Erroneous glucose recordings while using mutant variant of quinoprotein glucose dehydrogenase glucometer in a child with galactosemia.

Authors:  Vivek Mathew; Anantharaman Ramakrishnan; Ranjini Srinivasan; K Sushma; Ganapathi Bantwal; Vageesh Ayyar
Journal:  Indian J Endocrinol Metab       Date:  2013-10

4.  Influence of Vitamin C and Maltose on the Accuracy of Three Models of Glucose Meters.

Authors:  Jooyoung Cho; Sunyoung Ahn; Jisook Yim; Younjung Cheon; Seok Hoon Jeong; Sang Guk Lee; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2016-05       Impact factor: 3.464

Review 5.  Single Domain Antibodies as New Biomarker Detectors.

Authors:  Chiuan Herng Leow; Katja Fischer; Chiuan Yee Leow; Qin Cheng; Candy Chuah; James McCarthy
Journal:  Diagnostics (Basel)       Date:  2017-10-17
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.