Fundamental properties of the delivery of volumetric modulated arc therapy (VMAT) to static patient anatomy.

D Rangaraj1, S Oddiraju, B Sun, L Santanam, D Yang, S Goddu, L Papiez.   

Abstract

PURPOSE: The primary goal of this article is to formulate volumetric modulated arc therapy (VMAT) delivery problem and study interdependence between several parameters (beam dose rate, gantry angular speed, and MLC leaf speed) in the delivery of VMAT treatment plan. The secondary aim is to provide delivery solution and prove optimality (minimal beam on time) of the solution. An additional goal of this study is to investigate alternative delivery approaches to VMAT (like constant beam dose rate and constant gantry angular speed delivery).
METHOD: The problem of the VMAT delivery is formulated as a control problem with machine constraints. The relationships between parameters of arc therapy delivery are derived under the constraint of treatment plan invariance and limitations on delivery parameters. The nonuniqueness of arc therapy delivery solutions is revealed from these relations. The most efficient delivery of arc therapy is then formulated as optimal control problem and solved by geometrical methods. A computer program is developed to find numerical solutions for deliveries of specific VMAT plan.
RESULTS: Explicit examples of VMAT plan deliveries are computed and illustrated with graphical representations of the variability of delivery parameters. Comparison of delivery parameters with that of Varian's delivery are shown and discussed. Alternative delivery strategies such as constant gantry angular speed delivery and constant beam dose rate delivery are formulated and solutions are provided. The treatment times for all the delivery solutions are provided.
CONCLUSION: The investigations derive and prove time optimal VMAT deliveries. The relationships between delivery parameters are determined. The optimal alternative delivery strategies are discussed.

Entities:  

Mesh:

Year:  2010        PMID: 20879567     DOI: 10.1118/1.3453575

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Optimal partial-arcs in VMAT treatment planning.

Authors:  Jeremiah Wala; Ehsan Salari; Wei Chen; David Craft
Journal:  Phys Med Biol       Date:  2012-09-05       Impact factor: 3.609

2.  Searching standard parameters for volumetric modulated arc therapy (VMAT) of prostate cancer.

Authors:  Marius Treutwein; Matthias Hipp; Oliver Koelbl; Barbara Dobler
Journal:  Radiat Oncol       Date:  2012-07-11       Impact factor: 3.481

3.  Assessment of Volumetric-Modulated Arc Therapy for Constant and Variable Dose Rates.

Authors:  Mariluz De Ornelas-Couto; Ivaylo Mihaylov; Nesrin Dogan
Journal:  J Med Phys       Date:  2017 Oct-Dec

4.  Validation of the RapidArc Delivery System Using a Volumetric Phantom as Per Task Group Report 119 of the American Association of Physicists in Medicine.

Authors:  Lalit Kumar; Girigesh Yadav; Vimal Kishore; Manindra Bhushan; Munish Gairola; Deepak Tripathi
Journal:  J Med Phys       Date:  2019 Apr-Jun

5.  Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery.

Authors:  Soumya Roy; Biplab Sarkar; Anirudh Pradhan
Journal:  J Med Phys       Date:  2021-12-31

6.  Dosimetric and clinical outcomes after volumetric modulated arc therapy for carcinoma of the thoracic esophagus.

Authors:  Cai Xu; Mian Xi; Ritsuko Komaki; Peter A Balter; Meilin Huang; Brian P Hobbs; Luhua Wang; Steven H Lin
Journal:  Adv Radiat Oncol       Date:  2017-03-27