Literature DB >> 20879560

Design and optimization of large area thin-film CdTe detector for radiation therapy imaging applications.

E Ishmael Parsai1, Diana Shvydka, Jun Kang.   

Abstract

PURPOSE: The authors investigate performance of thin-film cadmium telluride (CdTe) in detecting high-energy (6 MV) x rays. The utilization of this material has become technologically feasible only in recent years due to significant development in large area photovoltaic applications.
METHODS: The CdTe film is combined with a metal plate, facilitating conversion of incoming photons into secondary electrons. The system modeling is based on the Monte Carlo simulations performed to determine the optimized CdTe layer thickness in combination with various converter materials.
RESULTS: The authors establish a range of optimal parameters producing the highest DQE due to energy absorption, as well as signal and noise spatial spreading. The authors also analyze the influence of the patient scatter on image formation for a set of detector configurations. The results of absorbed energy simulation are used in device operation modeling to predict the detector output signal. Finally, the authors verify modeling results experimentally for the lowest considered device thickness.
CONCLUSIONS: The proposed CdTe-based large area thin-film detector has a potential of becoming an efficient low-cost electronic portal imaging device for radiation therapy applications.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879560     DOI: 10.1118/1.3438082

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Low-cost flexible thin-film detector for medical dosimetry applications.

Authors:  P Zygmanski; C Abkai; Z Han; Y Shulevich; D Menichelli; J Hesser
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

2.  Radiation hardness of cadmium telluride solar cells in proton therapy beam mode.

Authors:  Shinhaeng Cho; Sang Hee Ahn; Ick Joon Cho; Yong Hyub Kim; Jae-Uk Jeong; Mee Sun Yoon; Sung-Ja Ahn; Woong-Ki Chung; Taek-Keun Nam; Ju-Young Song
Journal:  PLoS One       Date:  2019-09-12       Impact factor: 3.240

3.  Assessment of a Therapeutic X-ray Radiation Dose Measurement System Based on a Flexible Copper Indium Gallium Selenide Solar Cell.

Authors:  Dong-Seok Shin; Tae-Ho Kim; Jeong-Eun Rah; Dohyeon Kim; Hye Jeong Yang; Se Byeong Lee; Young Kyung Lim; Jonghwi Jeong; Haksoo Kim; Dongho Shin; Jaeman Son
Journal:  Sensors (Basel)       Date:  2022-08-04       Impact factor: 3.847

4.  Thin-film CdTe detector for microdosimetric study of radiation dose enhancement at gold-tissue interface.

Authors:  Nava Raj Paudel; Diana Shvydka; E Ishmael Parsai
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.