Literature DB >> 20879420

Calibration and use of intraoperative cone-beam computed tomography: an in-vitro study for wrist fracture.

Erin Janine Smith1, Anton Oentoro, Hisham Al-Sanawi, Braden Gammon, Paul St John, David R Pichora, Randy E Ellis.   

Abstract

The standard workflow in many image-guided procedures, preoperative imaging followed by intraoperative registration, can be a challenging process and is not readily adaptable to certain anatomical regions such as the wrist. In this study we present an alternative, consisting of a preoperative registration calibration and intraoperative navigation using 3D cone-beam CT. A custom calibration tool was developed to preoperatively register an optical tracking system to the imaging space of a digital angiographic C-arm. This preoperative registration was then applied to perform direct navigation using intraoperatively acquired images for the purposes of an in-vitro wrist fixation procedure. A validation study was performed to assess the stability of the registration and found that the mean registration error was approximately 0.3 mm. When compared to two conventional techniques, our navigated wrist repair achieved equal or better screw placement, with fewer drilling attempts and no additional radiation exposure to the patient. These studies suggest that preoperative registration coupled with direct navigation using procedure-specific graphical rendering, is potentially a highly accurate and effective means of performing image-guided interventions.

Entities:  

Mesh:

Year:  2010        PMID: 20879420     DOI: 10.1007/978-3-642-15711-0_45

Source DB:  PubMed          Journal:  Med Image Comput Comput Assist Interv


  5 in total

1.  Calibration of mobile-gantry computed tomography for surgical navigation.

Authors:  Anna Belkova; David R Pichora; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-10-08       Impact factor: 2.924

Review 2.  Computer-assisted Surgery for Scaphoid Fracture.

Authors:  Zi-Run Xiao; Ge Xiong
Journal:  Curr Med Sci       Date:  2018-12-07

3.  Volume slicing of cone-beam computed tomography images for navigation of percutaneous scaphoid fixation.

Authors:  Erin J Smith; Hisham A Al-Sanawi; Braden Gammon; Paul J St John; David R Pichora; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-06-25       Impact factor: 2.924

4.  Investigating the performance of a wrist stabilization device for image-guided percutaneous scaphoid fixation.

Authors:  Erin J Smith; Gregory Allan; Braden Gammon; Richard W Sellens; Randy E Ellis; David R Pichora
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-04-16       Impact factor: 2.924

5.  Computer-assisted percutaneous scaphoid fixation: concepts and evolution.

Authors:  Erin J Smith; Randy E Ellis; David R Pichora
Journal:  J Wrist Surg       Date:  2013-11
  5 in total