Numerically simulated cardiac exposure to electric current densities induced by TASER X-26 pulses in adult men.

N Leitgeb1, F Niedermayr, R Neubauer, G Loos.   

Abstract

There is still an ongoing debate whether or not electronic stun devices (ESDs) induce cardiac fibrillation. To assess the ventricular fibrillation risk of law enforcing electronic control devices, quantitative estimates of cardiac electric current densities induced by delivered electric pulses are essential. Numerical simulations were performed with the finite integration technique and the anatomical model of a standardized European man (NORMAN) segmented into 2 mm voxels and 35 different tissues. The load-dependent delivery of TASER X-26 pulses has been taken into account. Cardiac exposure to electric current densities of vertically and horizontally aligned dart electrodes was quantified and different hit scenarios compared. Since fibrillation thresholds critically depend on exposed volume, the provided quantitative data are essential for risk assessment. The maximum cardiac rms current densities amounted to 7730 A m(-2). Such high current densities and exposed cardiac volumes do not exclude ventricular fibrillation.

Entities:  

Mesh:

Year:  2010        PMID: 20876969     DOI: 10.1088/0031-9155/55/20/010

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

Review 1.  Cardiac and skeletal muscle effects of electrical weapons : A review of human and animal studies.

Authors:  Sebastian N Kunz; Hugh Calkins; Jiri Adamec; Mark W Kroll
Journal:  Forensic Sci Med Pathol       Date:  2018-06-28       Impact factor: 2.007

2.  Cardiac fibrillation risk of TASER X-26 dart mode application.

Authors:  Norbert Leitgeb; Florian Niedermayr; Gerhart Loos; Robert Neubauer
Journal:  Wien Med Wochenschr       Date:  2011-11-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.