Literature DB >> 20873706

Wavefront-guided photorefractive keratectomy with the VISX platform for myopia.

Simon R Bababeygy1, Edward E Manche.   

Abstract

PURPOSE: To evaluate the safety, efficacy, and clinical outcome of wavefront-guided photorefractive keratectomy (PRK) for the treatment of low to moderate and high myopia and compound myopic astigmatism.
METHODS: This retrospective study included 174 eyes of 102 patients with low to moderate (-0.125 to <-6.50 diopters [D]) and high (≥-6.50 D) myopia treated with wavefront-guided PRK using the VISX CustomVue S4 IR (Abbott Medical Optics) excimer laser. Eyes were analyzed preoperatively and at 1, 3, 6, and 12 months postoperatively. At 3 months, 174 eyes of 102 patients were evaluated, and at 12 months, 146 eyes of 75 patients were evaluated. No eyes were retreated in the study.
RESULTS: Mean patient age was 38.3±8.76 years (range: 20 to 66 years), and mean preoperative manifest refraction spherical equivalent (MRSE) was -5.60±2.58 D (range: -0.13 to -10.50 D). At 12 months postoperative, MRSE was -0.16±0.45 D (range: -1.50 to 1.38 D); 96.6% of eyes were within ±1.00 D of intended correction, whereas 81.5% were within ±0.50 D of intended correction. For all eyes, the safety and efficacy indexes were 1.01 and 1.02, respectively.
CONCLUSIONS: These data support the safety and efficacy for the correction of low to moderate and high myopia and compound myopic astigmatism using wavefront-guided PRK. Copyright 2011, SLACK Incorporated.

Entities:  

Mesh:

Year:  2010        PMID: 20873706     DOI: 10.3928/1081597X-20100527-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Wavefront-guided laser in situ keratomileusis (Lasik) versus wavefront-guided photorefractive keratectomy (Prk): a prospective randomized eye-to-eye comparison (an American Ophthalmological Society thesis).

Authors:  Edward E Manche; Weldon W Haw
Journal:  Trans Am Ophthalmol Soc       Date:  2011-12

2.  Prospective, randomized, fellow eye comparison of WaveLight Allegretto Wave Eye-Q versus VISX CustomVueTM STAR S4 IRTM in photorefractive keratectomy: analysis of visual outcomes and higher-order aberrations.

Authors:  Majid Moshirfar; Daniel S Churgin; Brent S Betts; Maylon Hsu; Shameema Sikder; Marcus Neuffer; Dane Church; Mark D Mifflin
Journal:  Clin Ophthalmol       Date:  2011-08-22

3.  Wavefront-Guided Photorefractive Keratectomy with the Use of a New Hartmann-Shack Aberrometer in Patients with Myopia and Compound Myopic Astigmatism.

Authors:  Steven C Schallhorn; Jan A Venter; Stephen J Hannan; Keith A Hettinger
Journal:  J Ophthalmol       Date:  2015-10-04       Impact factor: 1.909