Literature DB >> 20860242

Reducing errors during patient-controlled analgesia therapy through failure mode and effects analysis.

Janet Moss1.   

Abstract

BACKGROUND: Despite the technologic advances in design, resulting in the development of "smart" pumps to help deliver analgesia more safely, patient-controlled analgesia (PCA) is still involved in a significant proportion of the medication errors ascribed to intravenous (IV) drug administration, many of which have harmed patients. In 2003, Failure Mode and Effects Analysis (FMEA) was used to assess the PCA process at a 695-bed teaching and research tertiary hospital. IDENTIFYING AND ADDRESSING FAILURE MODES: For the three processes with hazard scores > 8--patient selection, prescribing, and medication administration-the potential cause(s) were identified, allowing the process to be redesigned to eliminate the potential cause(s).
RESULTS: In January 2003 to May 2003, before the FMEA process began, there were 11 PCA errors (extrapolated to 26 for the entire 2003 calendar year). In 2004, when most of the corrective actions were taken, there were 22 reported PCA errors. In October 2007, a new online occurrence-reporting program was implemented, making reporting much easier. From October 2007 through September 2008, there were only 8 reported PCA errors, representing a 69% reduction from baseline. No serious adverse events were associated with any of these PCA errors. DISCUSSION: Despite the reduction in PCA errors since the FMEA was conducted, misprogramming of drug concentration remains a common PCA error. Solutions include safety software for IV infusion pumps, an integral bar-code reader for detecting concentration errors, and interoperability of the software with other hospital information systems. One lesson learned was that an FMEA can lead to resolution of problems beyond the scope of original intent-in this case, the development of a new system for identifying all broken equipment.

Entities:  

Mesh:

Year:  2010        PMID: 20860242     DOI: 10.1016/s1553-7250(10)36054-5

Source DB:  PubMed          Journal:  Jt Comm J Qual Patient Saf        ISSN: 1553-7250


  5 in total

1.  Safeguarding the process of drug administration with an emphasis on electronic support tools.

Authors:  Hanna M Seidling; Anette Lampert; Kristina Lohmann; Julia T Schiele; Alexander J F Send; Diana Witticke; Walter E Haefeli
Journal:  Br J Clin Pharmacol       Date:  2013-09       Impact factor: 4.335

2.  Safety strategies in an academic radiation oncology department and recommendations for action.

Authors:  Stephanie A Terezakis; Peter Pronovost; Kendra Harris; Theodore Deweese; Eric Ford
Journal:  Jt Comm J Qual Patient Saf       Date:  2011-07

3.  The efficacy of elastomeric patient-control module when connected to a balloon pump for postoperative epidural analgesia: A randomized, noninferiority trial.

Authors:  Myung Hwa Kim; Yon Hee Shim; Min-Soo Kim; Yang-Sik Shin; Hyun Joo Lee; Jeong Soo Lee
Journal:  Medicine (Baltimore)       Date:  2017-01       Impact factor: 1.889

4.  Impact of Patient-Controlled Analgesia (PCA) Smart Pump-Electronic Health Record (EHR) Interoperability with Auto-Documentation on Chart Completion in a Community Hospital Setting.

Authors:  Tina M Suess; John W Beard; Barbara Trohimovich
Journal:  Pain Ther       Date:  2019-07-26

5.  Mapping outcomes in quality improvement and system design activities: the outcome identification loop and system impact model.

Authors:  Emmanuel Adeoluwa Akinluyi; Keith Ison; P John Clarkson
Journal:  BMJ Open Qual       Date:  2019-09-06
  5 in total