Long term performance characteristics of an electrochemical nitric oxide analyser.

D Robin Taylor1, Rochelle Palmay, Jan O Cowan, G Peter Herbison.   

Abstract

BACKGROUND: The NIOX MINO(®) is a nitric oxide (FE(NO)) analyser based on electrochemical technology. It includes a replaceable sensor. Quality control procedures are recommended, but regular calibration is not possible. We aimed to evaluate the performance characteristics of the NIOX MINO(®) to identify if reproducibility changed over time, or with different sensors. Also, there are reports that reproducibility of FE(NO) may be reduced in patients with high FE(NO): our secondary aim was to address this issue.
METHODS: Reproducibility in 24 separate sensor-analyser units was calculated on three occasions over two months in 17 patients. These included 9 patients whose FE(NO) was high (mean 80 ppb) and 8 in whom FE(NO) was low (mean 16 ppb).
RESULTS: One device failed quality control testing. For the remaining 23 sensor-analyser combinations, the mean coefficient of variation was 4.0% (range 1.2-7.2%) at baseline, 3.6% (range 2.0-7.0) at one month, and 3.6% (range 1.6-7.6%) at two months. The 95% C.I. for the mean limits of agreement for FE(NO) was ± 4.2 ppb (range 0.9-9.6 ppb), ± 3.8 ppb (range 1.6-6.9 ppb) and ± 3.2 ppb (range 1.2-6.8 ppb) respectively (NS). The limits of agreement exceeded the manufacturer's specifications (± 5 ppb) in 0 devices at baseline, 3 (13%) at one month, and 5 (22%) at two months.
CONCLUSIONS: Reproducibility of FE(NO) using the NIOX MINO(®) was within clinically acceptable limits (± 10 ppb) and was generally stable. However, with time, a proportion of individual sensor-analyser combinations yielded variability outside the manufacturer's specifications. Copyright Â
© 2010. Published by Elsevier Ltd.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20855188     DOI: 10.1016/j.rmed.2010.09.001

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  4 in total

Review 1.  An Evidence-Based Review of Application Devices for Nitric Oxide Concentration Determination from Exhaled Air in the Diagnosis of Inflammation and Treatment Monitoring.

Authors:  Magdalena Wyszyńska; Monika Nitsze-Wierzba; Aleksandra Czelakowska; Jacek Kasperski; Joanna Żywiec; Małgorzata Skucha-Nowak
Journal:  Molecules       Date:  2022-07-03       Impact factor: 4.927

2.  Impaired pulmonary nitric oxide bioavailability in pulmonary tuberculosis: association with disease severity and delayed mycobacterial clearance with treatment.

Authors:  Anna P Ralph; Tsin W Yeo; Cheryl M Salome; Govert Waramori; Gysje J Pontororing; Enny Kenangalem; Emiliana Tjitra; Richard Lumb; Graeme P Maguire; Ric N Price; Mark D Chatfield; Paul M Kelly; Nicholas M Anstey
Journal:  J Infect Dis       Date:  2013-06-03       Impact factor: 5.226

3.  Online Measurement of Exhaled NO Concentration and Its Production Sites by Fast Non-equilibrium Dilution Ion Mobility Spectrometry.

Authors:  Liying Peng; Dandan Jiang; Zhenxin Wang; Jiwei Liu; Haiyang Li
Journal:  Sci Rep       Date:  2016-03-15       Impact factor: 4.379

Review 4.  Fractional exhaled nitric oxide-measuring devices: technology update.

Authors:  Mauro Maniscalco; Carolina Vitale; Alessandro Vatrella; Antonio Molino; Andrea Bianco; Gennaro Mazzarella
Journal:  Med Devices (Auckl)       Date:  2016-06-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.