Literature DB >> 20851562

Safety analysis of all-inside arthroscopic repair of peripheral triangular fibrocartilage complex.

Scott M Waterman1, Dirk Slade, Brendan D Masini, Brett D Owens.   

Abstract

PURPOSE: The purpose of this study was to determine whether an all-inside peripheral triangular fibrocartilage complex (TFCC) repair using the FasT-Fix device (Smith & Nephew Endoscopy, Andover, MA) is safe by measuring the proximity of the anchors to ulnar-sided anatomic structures.
METHODS: Eleven fresh-frozen cadaveric wrists were thawed and placed in traction. Under direct arthroscopic visualization, an all-inside arthroscopic peripheral TFCC repair was completed by placing a single FasT-Fix device in a vertical mattress fashion. The wrists were then dissected to visualize the 2 anchors. The distance between these anchors and the flexor carpi ulnaris (FCU), extensor carpi ulnaris (ECU), and dorsal branch of the ulnar sensory nerve (DBUN) were measured with digital calipers and recorded.
RESULTS: The peripheral anchor averaged 4.2 mm (range, 0 to 14 mm) from the ECU tendon, 3.8 mm (range, 0 to 9 mm) from the DBUN, and 8.3 mm (range, 1 to 15 mm) from the FCU tendon. The central anchor averaged 9.6 mm (range, 2 to 15 mm) from the ECU tendon, 6.8 mm (range, 1 to 13 mm) from the DBUN, and 7.6 mm (range, 1 to 13 mm) from the FCU tendon.
CONCLUSIONS: This study exposes some safety concerns with the all-inside peripheral TFCC repair using the FasT-Fix device, which was found to reside in close proximity to the ECU, FCU, and DBUN. In multiple wrists the anchors were noted to underlie the anatomic structure that we measured, making it possible to pierce these structures with the needle before deployment of the anchor. CLINICAL RELEVANCE: Though technically feasible, all-inside arthroscopic repair of the peripheral TFCC risks injury to the ulnar-sided anatomy. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20851562     DOI: 10.1016/j.arthro.2010.02.027

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  2 in total

1.  Long-Term Outcomes of All-Arthroscopic Pre-Tied Suture Device Triangular Fibrocartilage Complex Repair.

Authors:  Eric J Sarkissian; Matthew B Burn; Jeffrey Yao
Journal:  J Wrist Surg       Date:  2019-05-27

2.  Neurologic complications in common wrist and hand surgical procedures.

Authors:  Nicole Verdecchia; Julie Johnson; Mark Baratz; Steven Orebaugh
Journal:  Orthop Rev (Pavia)       Date:  2018-03-29
  2 in total