| Literature DB >> 20850961 |
Md Shahinul Islam1, Hyunung Yu, Hee Gu Lee, Seong Ho Kang.
Abstract
A novel detection technique on biochip for the quantification of label-free C-reactive protein (CRP) based on molecular switching of fluorescence (MSF) is demonstrated by total internal reflection fluorescence microscopy. It alters fluorescence intensity of fluoreseinamine isomer 1 (FAI) upon binding with its specific ligand, O-phosphorylethanolamine (PEA). In the MSF-based detection, FAI was used as an ink, printed on a 3-glycidoxypropyl-trimethoxysilane (GPTS)-coated glass coverslip. With the addition of GPTS conjugated PEA solution to the FAI-printed coverslip, the fluorescence intensity was remarkably decreased. Addition of CRP increased fluorescence intensity linearly in the range of 800 aM to 500 fM (R=0.997). The MSF-based biochip assay for the estimation of CRP in human sera showed ∼200 times increased detection sensitivity in less than a third of the time to obtain results using a conventional enzyme-linked immunosorbent assay. This biochip detection is a promising new technique for the quantification of CRP molecules from trace amounts of clinical samples.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20850961 DOI: 10.1016/j.bios.2010.08.045
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618