Literature DB >> 20841783

Enhanced notification of infusion pump programming errors.

R Scott Evans1, Rick Carlson, Kyle V Johnson, Brent K Palmer, James F Lloyd.   

Abstract

Hospitalized patients receive countless doses of medications through manually programmed infusion pumps. Many medication errors are the result of programming incorrect pump settings. When used appropriately, smart pumps have the potential to detect some programming errors. However, based on the current use of smart pumps, there are conflicting reports on their ability to prevent patient harm without additional capabilities and interfaces to electronic medical records (EMR). We developed a smart system that is connected to the EMR including medication charting that can detect and alert on potential pump programming errors. Acceptable programming limits of dose rate increases in addition to initial drug doses for 23 high-risk medications are monitored. During 22.5 months in a 24 bed ICU, 970 alerts (4% of 25,040 doses, 1.4 alerts per day) were generated for pump settings programmed outside acceptable limits of which 137 (14%) were found to have prevented potential harm. Monitoring pump programming at the system level rather than the pump provides access to additional patient data in the EMR including previous dosage levels, other concurrent medications and caloric intake, age, gender, vitals and laboratory results.

Entities:  

Mesh:

Year:  2010        PMID: 20841783

Source DB:  PubMed          Journal:  Stud Health Technol Inform        ISSN: 0926-9630


  7 in total

1.  Safeguarding the process of drug administration with an emphasis on electronic support tools.

Authors:  Hanna M Seidling; Anette Lampert; Kristina Lohmann; Julia T Schiele; Alexander J F Send; Diana Witticke; Walter E Haefeli
Journal:  Br J Clin Pharmacol       Date:  2013-09       Impact factor: 4.335

Review 2.  Benefits and risks of using smart pumps to reduce medication error rates: a systematic review.

Authors:  Kumiko Ohashi; Olivia Dalleur; Patricia C Dykes; David W Bates
Journal:  Drug Saf       Date:  2014-12       Impact factor: 5.606

Review 3.  Electronic Health Records: Then, Now, and in the Future.

Authors:  R S Evans
Journal:  Yearb Med Inform       Date:  2016-05-20

4.  The urgent need for innovation in I.V. smart pumps.

Authors:  Karen K Giuliano; Charles Niemi
Journal:  Nurs Manage       Date:  2015-03

Review 5.  The urgent need for innovation in I.V. infusion devices.

Authors:  Karen K Giuliano; Charles Niemi
Journal:  Nursing       Date:  2016-04

6.  Difficulty Using Smart Pump Logs to Recreate a Patient Safety Event: Case Study and Considerations for Pump Enhancements.

Authors:  Andrew A M Ibey; Derek Andrews; Barb Ferreira
Journal:  Drug Saf Case Rep       Date:  2016-12

7.  Systemic Defenses to Prevent Intravenous Medication Errors in Hospitals: A Systematic Review.

Authors:  Sini Karoliina Kuitunen; Ilona Niittynen; Marja Airaksinen; Anna-Riia Holmström
Journal:  J Patient Saf       Date:  2021-12-01       Impact factor: 2.243

  7 in total