Evaluation of EBT-2 model GAFCHROMIC film performance in water.

Saad Aldelaijan1, Slobodan Devic, Huriyyah Mohammed, Nada Tomic, Li-Heng Liang, Francois DeBlois, Jan Seuntjens.   

Abstract

PURPOSE: The authors present results of the measurements on the impact of radiochromic film immersion in water. The impact of film piece size, initial optical density, postimmersion waiting time prior to scanning, and the time film was kept in water has been investigated. The authors also investigated the pathways of water penetration into the film during the film immersion in water.
METHODS: To study the impact of water immersion on change in optical density, the authors used various sizes of the latest EBT-2 model GAFCHROMICTM film: 2 x 2, 4 x 4, and 8 x 8 in.2. In addition, to test any existing dependence of the film's optical density on water diffusion, the authors used two sets of films: Unexposed (0 Gy) and film pieces exposed to a dose of 3 Gy. Times that film pieces were left in water ranged from 30 min to 24 h, and once the film was permanently removed from water, the authors also studied the impact of the scanning time (deltat) that ranged from 0 (films scanned right after removal from water) to 72 h postimmersion.
RESULTS: While the penetration depth can reach as much as 9 mm around the edges of the EBT-2 GAFCHROMIC film, the anticipated dose error due to the change in optical density due to the water immersion appears to be negligible for the short immersions of the order of 30 min. However, as the immersion time increases, the anticipated dose error may reach 22 cGy on a 2 x 2 in.2 piece of film, which corresponds to 7% dose error at 3 Gy of measured dose.
CONCLUSIONS: In this work, the authors report on an undoubted impact of radiochromic film immersion in water on the measured change in optical density, which may lead to systematic errors in dose measurements if the film is kept in water for longer periods of time. The magnitude of the impact depends on many parameters: Size of the film piece, initial optical density, postimmersion waiting time prior to scanning (defined by the current radiochromic film dosimetry protocol in. place), and the time film was kept in water. The authors also suggested various approaches in correcting for the change in netOD due to water penetration into the film, but the authors believe that the use of the control film piece would be the most appropriate.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20831076     DOI: 10.1118/1.3455713

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Comparison of trigeminal neuralgia radiosurgery plans using two film detectors for the commissioning of small photon beams.

Authors:  Karina P Esparza-Moreno; Olivia A García-Garduño; Paola Ballesteros-Zebadúa; José M Lárraga-Gutiérrez; Sergio Moreno-Jiménez; Miguel A Celis-Lopez
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

2.  Investigation of EBT3 radiochromic film's response to humidity.

Authors:  Elsa Y León-Marroquín; José M Lárraga-Gutiérrez; J Alfredo Herrera-González; Miguel A Camacho-López; José E Villarreal Barajas; Olivia A García-Garduño
Journal:  J Appl Clin Med Phys       Date:  2018-04-29       Impact factor: 2.102

3.  Characterization of GafChromic EBT2 film dose measurements using a tissue-equivalent water phantom for a Theratron® Equinox Cobalt-60 teletherapy machine.

Authors:  Daniel Akwei Addo; Elsie Effah Kaufmann; Samuel Nii Tagoe; Augustine Kwame Kyere
Journal:  PLoS One       Date:  2022-08-19       Impact factor: 3.752

4.  Radiochromic film based dosimetry of image-guidance procedures on different radiotherapy modalities.

Authors:  Ahmad Nobah; Saad Aldelaijan; Slobodan Devic; Nada Tomic; Jan Seuntjens; Mohammed Al-Shabanah; Belal Moftah
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

5.  Investigation of a source model for a new electronic brachytherapy tandem by film measurement.

Authors:  Elijah Martin; Keith Sowards; Brian Wang
Journal:  J Appl Clin Med Phys       Date:  2018-08-13       Impact factor: 2.102

  5 in total