Literature DB >> 20831061

Direct absorbed dose to water determination based on water calorimetry in scanning proton beam delivery.

A Sarfehnia1, B Clasie, E Chung, H M Lu, J Flanz, E Cascio, M Engelsman, H Paganetti, J Seuntjens.   

Abstract

PURPOSE: The aim of this manuscript is to describe the direct measurement of absolute absorbed dose to water in a scanned proton radiotherapy beam using a water calorimeter primary standard.
METHODS: The McGill water calorimeter, which has been validated in photon and electron beams as well as in HDR 192Ir brachytherapy, was used to measure the absorbed dose to water in double scattering and scanning proton irradiations. The measurements were made at the Massachusetts General Hospital proton radiotherapy facility. The correction factors in water calorimetry were numerically calculated and various parameters affecting their magnitude and uncertainty were studied. The absorbed dose to water was compared to that obtained using an Exradin T1 Chamber based on the IAEA TRS-398 protocol.
RESULTS: The overall 1-sigma uncertainty on absorbed dose to water amounts to 0.4% and 0.6% in scattered and scanned proton water calorimetry, respectively. This compares to an overall uncertainty of 1.9% for currently accepted IAEA TRS-398 reference absorbed dose measurement protocol. The absorbed dose from water calorimetry agrees with the results from TRS-398 well to within 1-sigma uncertainty.
CONCLUSIONS: This work demonstrates that a primary absorbed dose standard based on water calorimetry is feasible in scattered and scanned proton beams.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20831061     DOI: 10.1118/1.3427317

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Ultrahigh dose rate pencil beam scanning proton dosimetry using ion chambers and a calorimeter in support of first in-human FLASH clinical trial.

Authors:  Eunsin Lee; Ana Mónica Lourenço; Joseph Speth; Nigel Lee; Anna Subiel; Francesco Romano; Russell Thomas; Richard A Amos; Yongbin Zhang; Zhiyan Xiao; Anthony Mascia
Journal:  Med Phys       Date:  2022-07-14       Impact factor: 4.506

  1 in total