Wide field array calibration dependence on the stability of measured dose distributions.

Thomas A Simon1, William E Simon, Darren Kahler, Jonathan Li, Chihray Liu.   

Abstract

PURPOSE: The aim of this work was to simulate the effect of dose distribution changes on detector array calibrations and to explore compensatory methods that are used during calibration measurements.
METHODS: The array calibration technique that was investigated is known as wide field (WF) calibration. Using this method, a linear array [y-axis (65 detectors) of the IC PROFILER (Sun Nuclear Corporation, Melbourne, FL)] is calibrated with three measurements (alpha, theta, and lamda); each measurement uses the same radiation field, which is larger than the array. For measurement configuration theta, the array is rotated by 180 degrees from its position in a; for lamda, the array is shifted by one detector from its position in theta. The relative detector sensitivities are then determined through ratios of detector readings at the same field locations (using theta and lamda). This method results in error propagation that is proportional to the number of detectors in the array. During the procedure, the calibration protocol operates under three postulates, which state that (a) the beam shape does not change between measurements; (b) the relative sensitivities of the detectors do not change; and (c) the scatter to the array does not change as the array is moved. The WF calibration's sensitivity to a postulate (a) violation was quantified by applying a sine shaped perturbation (of up to 0.1%) to a, theta, or lamda, and then determining the change relative to a baseline calibration. Postulate (a) violations were minimized by using a continuous beam and mechanized array movement during theta and lamda. A continuously on beam demonstrated more stable beam symmetry as compared to cycling the beam on and off between measurements. Additional side-scatter was also used to satisfy postulate (c).
RESULTS: Simulated symmetry perturbations of 0.1% to theta or lamda resulted in calibration errors of up to 2%; alpha was relatively immune to perturbation (<0.1% error). Wide field calibration error on a linear accelerator with similar symmetry variations was +/- 1.6%. Using a continuous beam during theta and lamda with additional side-scatter reduced the calibration error from +/- 1.6% to +/- 0.48%.
CONCLUSIONS: This work increased the reproducibility of WF calibrations by limiting the effect of measurement perturbations primarily due to linear accelerator symmetry variations. The same technique would work for any array using WF calibration.

Mesh:

Year:  2010        PMID: 20831056     DOI: 10.1118/1.3442028

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID).

Authors:  Alberto Boriano; Francesco Lucio; Elisa Calamia; Elvio Russi; Flavio Marchetto
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

2.  A comparison of methods for monitoring photon beam energy constancy.

Authors:  Song Gao; Peter A Balter; Mark Rose; William E Simon
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

3.  Clinical implementation of photon beam flatness measurements to verify beam quality.

Authors:  Simon Goodall; Nicholas Harding; Jake Simpson; Louise Alexander; Steve Morgan
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

4.  Normalize the response of EPID in pursuit of linear accelerator dosimetry standardization.

Authors:  Bin Cai; S Murty Goddu; Sridhar Yaddanapudi; Douglas Caruthers; Jie Wen; Camille Noel; Sasa Mutic; Baozhou Sun
Journal:  J Appl Clin Med Phys       Date:  2017-11-10       Impact factor: 2.102

5.  Quantification of beam steering with an ionization chamber array.

Authors:  Song Gao; Peter A Balter; Benjamin Tran; Mark Rose; William E Simon
Journal:  J Appl Clin Med Phys       Date:  2018-03-25       Impact factor: 2.102

6.  Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.

Authors:  Saeed Ahmed; Geoffrey Zhang; Eduardo G Moros; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2019-09-03       Impact factor: 2.102

7.  Determination of the electronic portal imaging device pixel-sensitivity-map for quality assurance applications. Part 1: Comparison of methods.

Authors:  Michael Paul Barnes; Baozhou Sun; Brad Michael Oborn; Bishnu Lamichhane; Stuart Szwec; Matthew Schmidt; Bin Cai; Frederick Menk; Peter Greer
Journal:  J Appl Clin Med Phys       Date:  2022-04-15       Impact factor: 2.243

8.  Calibration of a detector array through beam profile reconstruction with error-locking.

Authors:  Song Wang; Zhiqiu Li; K S Clifford Chao; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.