Dosimetric verification in inhomogeneous phantom geometries for the XiO radiotherapy treatment planning system with 6-MV photon beams.

Ryosuke Kohno1, Satoshi Kitou, Eriko Hirano, Satoru Kameoka, Tomonori Goka, Teiji Nishio, Tomoko Miyagishi, Takaki Ariji, Mitsuhiko Kawashima, Takashi Ogino.   

Abstract

We have developed a practical dose verification method for radiotherapy treatment planning systems by using only a Farmer ionization chamber in inhomogeneous phantoms. In particular, we compared experimental dose verifications of multi-layer phantom geometries and laterally inhomogeneous phantom geometries for homogeneous and inhomogenous dose calculations by using the fast-Fourier-transform convolution, fast-superposition, and superposition in the XiO radiotherapy treatment-planning system. We applied the dose verification method to three kernel-based algorithms in various phantom geometries with water-, lung- and bone-equivalent media of different field sizes. These calculations were then compared with experimental measurements by use of the Farmer ionization chamber. The fast-Fourier-transform convolution algorithm overestimated the dose by about 8% in the lung phantom geometry. The superposition algorithm and the fast-superposition algorithm were both accurate to better than 2% when compared to the measurements even for complex geometries. Our dose verification method was able to clarify the differences and equivalences of the three kernel-based algorithms and measurements with use only of commonly available apparatus. This will be generally useful in commissioning of inhomogeneity-correction algorithms in the clinical practice of treatment planning.

Entities:  

Mesh:

Year:  2008        PMID: 20821134     DOI: 10.1007/s12194-008-0049-7

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  19 in total

1.  Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy.

Authors:  Nesrin Dogan; Leonid B Leybovich; Anil Sethi
Journal:  Phys Med Biol       Date:  2002-11-21       Impact factor: 3.609

2.  Dosimetric evaluation of a MOSFET detector for clinical application in photon therapy.

Authors:  Ryosuke Kohno; Eriko Hirano; Teiji Nishio; Tomoko Miyagishi; Tomonori Goka; Mitsuhiko Kawashima; Takashi Ogino
Journal:  Radiol Phys Technol       Date:  2007-11-17

3.  Comparison of dose calculation algorithms in slab phantoms with cortical bone equivalent heterogeneities.

Authors:  P Carrasco; N Jornet; M A Duch; V Panettieri; L Weber; T Eudaldo; M Ginjaume; M Ribas
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

4.  Generation of photon energy deposition kernels using the EGS Monte Carlo code.

Authors:  T R Mackie; A F Bielajew; D W Rogers; J J Battista
Journal:  Phys Med Biol       Date:  1988-01       Impact factor: 3.609

5.  Fast Fourier transform convolution calculations of x-ray isodose distributions in homogeneous media.

Authors:  A L Boyer; Y P Zhu; L Wang; P Francois
Journal:  Med Phys       Date:  1989 Mar-Apr       Impact factor: 4.071

6.  Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media.

Authors:  A Ahnesjö
Journal:  Med Phys       Date:  1989 Jul-Aug       Impact factor: 4.071

7.  Comments on dose measurements for a narrow beam in radiosurgery.

Authors:  A Wu; R D Zwicker; A M Kalend; Z Zheng
Journal:  Med Phys       Date:  1993 May-Jun       Impact factor: 4.071

8.  Commissioning and quality assurance of treatment planning computers.

Authors:  J Van Dyk; R B Barnett; J E Cygler; P C Shragge
Journal:  Int J Radiat Oncol Biol Phys       Date:  1993-05-20       Impact factor: 7.038

9.  Experimental validation tests of fast Fourier transform convolution and multigrid superposition algorithms for dose calculation in low-density media.

Authors:  Feliciano García-Vicente; Angel Miñambres; Inmaculada Jerez; Ignaci Modolell; Leopoldo Pérez; Juan Jose Torres
Journal:  Radiother Oncol       Date:  2003-05       Impact factor: 6.280

10.  Comparison of RTP dose distributions in heterogeneous phantoms with the BEAM Monte Carlo simulation system.

Authors:  M Miften; M Wiesmeyer; A Kapur; C M Ma
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

View more
  1 in total

1.  Evaluation of the usefulness of a MOSFET detector in an anthropomorphic phantom for 6-MV photon beam.

Authors:  Ryosuke Kohno; Eriko Hirano; Satoshi Kitou; Tomonori Goka; Kana Matsubara; Satoru Kameoka; Taeko Matsuura; Takaki Ariji; Teiji Nishio; Mitsuhiko Kawashima; Takashi Ogino
Journal:  Radiol Phys Technol       Date:  2010-02-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.