Optical coherence tomography assessment of a coronary bare cobalt chromium stent deformed by the removal of an entrapped "jailed" guidewire.

Pierfrancesco Agostoni1, Eric Van Belle, Pieter R Stella.   

Abstract

A 64-year-old male with an ongoing acute coronary syndrome was treated percutaneously for a subtotal occlusion of a first diagonal branch. After predilatation of the diagonal, an important secondary branch was evident at the level of the subocclusion. A jailed wire was placed in this vessel, and stenting of the main diagonal branch was performed with a last-generation thin-strut bare cobalt-chromium stent. Conventional retrieval of the jailed wire was impossible, as the distal tip of the wire remained entrapped at the level of the stent. The strong pulling of the wire led only to a deep intubation of the guiding catheter with possible risk of dissection. Retrieval of the wire was finally possible with the support of an over-the-wire microcatheter. However, the stent, though still patent, appeared deformed. At a scheduled control angiogram 4 months later (while the patient was still asymptomatic), an optical coherence tomography (OCT) investigation of the stented segment confirmed the deformation of the stent. OCT showed a complete lack of stent struts on the side of origin of the secondary branch and an "overlap" or "accumulation" of several layers of stent struts on the other side. There were no signs of uncovered struts overall nor malapposition, and even the sites where multiple stent layers were visible were completely covered by neointimal hyperplasia. Despite the multiple layers of metal, the neointimal reaction was moderate and did not lead to a flow-limiting stenosis, thus the patient was further treated medically without the need for a new intervention.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20814056

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  1 in total

1.  Longitudinal Stent Deformation Caused by Retraction of the Looped Main Branch Guidewire.

Authors:  Hung-Hao Lee; Po-Chao Hsu; Wen-Hsien Lee; Chun-Yuan Chu; Ho-Ming Su; Tsung-Hsien Lin; Wen-Chol Voon; Wen-Ter Lai; Sheng-Hsiung Sheu; Cheng-An Chiu
Journal:  Acta Cardiol Sin       Date:  2016-09       Impact factor: 2.672

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.