A comparison of knot security and loop security in arthroscopic knots tied with newer high-strength suture materials.

Ian K Y Lo1, Eloy Ochoa, Stephen S Burkhart.   

Abstract

PURPOSE: The purpose of this study was to compare the knot security and loop security of 2 sliding and 1 static arthroscopic knot tied with different types of suture material.
METHODS: We evaluated 3 commonly used arthroscopic knots (surgeon's knot, Roeder knot, and Weston knot) tied with 6 different braided No. 2 sutures (FiberWire [Arthrex, Naples, FL]; Ethibond [Ethicon, Somerville, NJ]; Orthocord [DePuy Mitek, Raynham, MA]; Herculine [now called HiFi; ConMed Linvatec, Largo, FL]; MaxBraid [Arthrotek, Warsaw, IN]; and UltraBraid [Smith & Nephew, Andover, MA]). Each suture loop was then mounted on a materials testing system, and its circumference was measured at a 5-N preload to assess each knot's ability to maintain a tight suture loop without slippage (loop security). Knot security was measured as the maximum force to failure at 3 mm of crosshead displacement or suture breakage during single-pull load testing.
RESULTS: We found that tying knots with different types of suture material can affect both the knot security and loop security of various types of arthroscopic knots. When a Roeder knot or surgeon's knot was tied, No. 2 FiberWire had the highest force to failure when compared with similar knots tied with other suture material (P < .001). The loop security for many of the knot and suture configurations was not significantly different. However, No. 2 FiberWire consistently showed the smallest loop circumference when compared with other suture materials.
CONCLUSIONS: Arthroscopic knots tied with No. 2 FiberWire provide superior knot security and similar loop security compared with other commonly used high-strength polyethylene suture materials. CLINICAL RELEVANCE: High-strength sutures exhibit unique mechanical characteristics that may vary significantly between suture types. In addition, knot configuration plays an important role in altering these characteristics as they relate to knot security.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20810086     DOI: 10.1016/j.arthro.2009.12.009

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  14 in total

1.  Beyond the square knot: a novel knotting technique for surgical use.

Authors:  Chunfeng Zhao; Chung-Chen Hsu; Tamami Moriya; Andrew R Thoreson; Steven S Cha; Steven L Moran; Kai-Nan An; Peter C Amadio
Journal:  J Bone Joint Surg Am       Date:  2013-06-05       Impact factor: 5.284

2.  Mechanical testing of different knot types using high-performance suture material.

Authors:  M H Baums; Ch Sachs; T Kostuj; K Schmidt-Horlohé; W Schultz; H-M Klinger
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-10-13       Impact factor: 4.342

3.  Effects of using a surgical clamp to hold tension while tying knots with commonly used orthopedic sutures.

Authors:  Leslie J Bisson; Andrew D Sobel; David Godfrey
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-09-16       Impact factor: 4.342

4.  A Biomechanical Comparison of Varying Base Knot Configurations with Different Overhand/Underhand Combinations of Reversing Half-Hitches on Alternating Posts After Basic Instructional Training.

Authors:  Heather A Evin; Tyler T Bilden; Benjamin C Noonan; Alexander Cm Chong
Journal:  Iowa Orthop J       Date:  2019

5.  Construct validity testing of the Arthroscopic Knot Trainer (ArK).

Authors:  Ivan H Wong; Matthew Denkers; Nathan Urquhart; Forough Farrokhyar
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-05-18       Impact factor: 4.342

6.  Optimization of the racking hitch knot: how many half hitches and which suture material provide the greatest security?

Authors:  James D Kelly; Suketu Vaishnav; Bradley M Saunders; Mark A Schrumpf
Journal:  Clin Orthop Relat Res       Date:  2014-02-11       Impact factor: 4.176

7.  Comparison of knot-tying proficiency and knot characteristics for square and reversing half hitch alternating-post surgical knots in a simulated deep body cavity among notice medical students.

Authors:  Vincent Wu; Cynthia Yeung; Edward A Sykes; Boris Zevin
Journal:  Can J Surg       Date:  2018-12-01       Impact factor: 2.089

8.  A biomechanical study of the medial row tightening on stability of single stitch suture-bridge construct.

Authors:  Sharon Abihssira; Pierre-Alban Bouche; Claire Cammas; Clément Thirache; Geoffroy Nourissat
Journal:  J Orthop Surg Res       Date:  2022-07-15       Impact factor: 2.677

9.  Loop and knot security of a novel arthroscopic sliding-locking knot using high-strength sutures.

Authors:  Cpt Sally Corey; Karl Wenger; Cpt Peter Johnson; Maj Stephen Parada
Journal:  J Orthop       Date:  2018-08-24

10.  Comparing the tensile strength of square and reversing half-hitch alternating post knots.

Authors:  Vincent Wu; Edward A Sykes; Dale Mercer; Wilma M Hopman; Ephraim Tang
Journal:  Can J Surg       Date:  2017-06       Impact factor: 2.089

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.