Simplified method for international normalized ratio (INR) derivation based on the prothrombin time/INR line: an international study.

Leon Poller1, Saied Ibrahim, Michelle Keown, Albert Pattison, Jørgen Jespersen.   

Abstract

BACKGROUND: The need to perform local International Sensitivity Index (ISI) calibrations and in particular the requirement for a manual method for prothrombin time (PT) determination, have proved to be obstacles to application of the WHO scheme for PT standardization.
METHODS: We used international normalized ratio (INR) derived with a set of only 5 European Concerted Action on Anticoagulation (ECAA) lyophilized calibrant plasmas, certified manually by expert centers with reference thromboplastins, to determine a local PT/INR Line. We compared results of an independent set of validation plasmas with INRs from conventional ISI calibrations and with manually certified INRs.
RESULTS: The mean certified INR of 5 lyophilized validation plasmas was 2.41 with human thromboplastin, 2.04 with bovine/combined, and 2.80 with rabbit. With 42 human reagents, the mean observed INR of the validation plasmas was 2.68 (11.2% deviation from certified INR). Deviation was reduced to 0.4% with both local ISI calibration and the PT/INR Line. Eight results based on bovine/combined thromboplastin gave an INR deviation of 4.9%, becoming 0.5% after ISI calibration and 2.4% with the PT/INR Line. Six results with rabbit reagents deviated from certified INR by 2.5%. After ISI calibration, deviation became 1.1%, and with the PT/INR Line, 0.7%. The PT/INR Line gave similar results with both linear and orthogonal regression analysis. The total proportion of validation plasmas giving INR within 10% deviation from certified values was 42.5% with uncorrected INR, which increased to 92.1% with local ISI calibration and 93.2% with the PT/INR Line.
CONCLUSIONS: The PT/INR Line procedure with 5 ECAA calibrant plasmas successfully substitutes for local ISI calibrations in deriving reliable INRs.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20798354     DOI: 10.1373/clinchem.2009.141937

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  Uncertainty of Measurement: A Review of the Rules for Calculating Uncertainty Components through Functional Relationships.

Authors:  Ian Farrance; Robert Frenkel
Journal:  Clin Biochem Rev       Date:  2012-05

2.  The need for an adapted initiation nomogram during Fiix prothrombin time monitoring of warfarin.

Authors:  Pall T Onundarson; Brynja R Gudmundsdottir
Journal:  J Thromb Thrombolysis       Date:  2019-11       Impact factor: 2.300

3.  Canalization effect in the coagulation cascade and the interindividual variability of oral anticoagulant response. A simulation study.

Authors:  Alexandru D Corlan; John Ross
Journal:  Theor Biol Med Model       Date:  2011-10-12       Impact factor: 2.432

4.  Data documenting the performance of the PT/INR line correction method for reconciling INR discrepancies between central laboratory coagulation analyzers using different thromboplastins during the evaluation of a portable Coagulometer.

Authors:  Wendy S Baker; Kathleen J Albright; Heidi Spratt; Megan Berman; Peggy A Mann; Jaime Unabia; John R Petersen
Journal:  Data Brief       Date:  2017-11-07