Literature DB >> 20796200

Towards the development of a rapid, portable, surface enhanced Raman spectroscopy based cleaning verification system for the drug nelarabine.

Damion K Corrigan1, Neale A Salton, Chris Preston, Sergey Piletsky.   

Abstract

OBJECTIVES: Cleaning verification is a scientific and economic problem for the pharmaceutical industry. A large amount of potential manufacturing time is lost to the process of cleaning verification. This involves the analysis of residues on spoiled manufacturing equipment, with high-performance liquid chromatography (HPLC) being the predominantly employed analytical technique. The aim of this study was to develop a portable cleaning verification system for nelarabine using surface enhanced Raman spectroscopy (SERS).
METHODS: SERS was conducted using a portable Raman spectrometer and a commercially available SERS substrate to develop a rapid and portable cleaning verification system for nelarabine. Samples of standard solutions and swab extracts were deposited onto the SERS active surfaces, allowed to dry and then subjected to spectroscopic analysis. KEY
FINDINGS: Nelarabine was amenable to analysis by SERS and the necessary levels of sensitivity were achievable. It is possible to use this technology for a semi-quantitative limits test. Replicate precision, however, was poor due to the heterogeneous drying pattern of nelarabine on the SERS active surface. Understanding and improving the drying process in order to produce a consistent SERS signal for quantitative analysis is desirable.
CONCLUSIONS: This work shows the potential application of SERS for cleaning verification analysis. SERS may not replace HPLC as the definitive analytical technique, but it could be used in conjunction with HPLC so that swabbing is only carried out once the portable SERS equipment has demonstrated that the manufacturing equipment is below the threshold contamination level.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20796200     DOI: 10.1111/j.2042-7158.2010.01152.x

Source DB:  PubMed          Journal:  J Pharm Pharmacol        ISSN: 0022-3573            Impact factor:   3.765


  3 in total

1.  Field detection devices for screening the quality of medicines: a systematic review.

Authors:  Serena Vickers; Matthew Bernier; Stephen Zambrzycki; Facundo M Fernandez; Paul N Newton; Céline Caillet
Journal:  BMJ Glob Health       Date:  2018-08-29

Review 2.  Recent advances in biomedical, biosensor and clinical measurement devices for use in humans and the potential application of these technologies for the study of physiology and disease in wild animals.

Authors:  Alexander Macdonald; Lucy A Hawkes; Damion K Corrigan
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  2021-06-28       Impact factor: 6.671

3.  A Microelectrode Array with Reproducible Performance Shows Loss of Consistency Following Functionalization with a Self-Assembled 6-Mercapto-1-hexanol Layer.

Authors:  Damion K Corrigan; Vincent Vezza; Holger Schulze; Till T Bachmann; Andrew R Mount; Anthony J Walton; Jonathan G Terry
Journal:  Sensors (Basel)       Date:  2018-06-09       Impact factor: 3.576

  3 in total