| Literature DB >> 2079057 |
B Frenette1, D Mergler, J Ferraris.
Abstract
The objective of the present study was to evaluate the reliability of a modified version of the commercially available Biothesiometer, and to examine vibrotactile perception thresholds with respect to age and gender. A standardized protocol for measuring vibrotactile perception threshold was administered to 80 subjects, once a week over 4 weeks. Inter-session variability was stable (analysis of variance for repeated measures; P greater than 0.05) and correlations were high (Pearson's: 0.87 less than or equal to r less than or equal to 0.90; P less than or equal to 0.001). For sites on both hands and feet, there was a significant increase with age (0.19 less than or equal to r2 less than or equal to 0.52; P less than or equal to 0.001). Five factor analysis of variance model showed that vibrotactile perception threshold was significantly different with stimulus site, age category and gender; no differences were observed with alcohol consumption or smoking status. The findings indicate that the measurements from this device are highly reproducible and sensitive to expected threshold differences with age and gender. The authors attribute this to technical improvements of the original apparatus, rigid adherence to test protocol and maintenance of standard conditions. This type of instrument would be useful in assessing vibrotactile perception loss in occupational health studies.Entities:
Mesh:
Year: 1990 PMID: 2079057 DOI: 10.1007/BF00236057
Source DB: PubMed Journal: Eur J Appl Physiol Occup Physiol ISSN: 0301-5548