| Literature DB >> 2074412 |
A Katz1, I Kaffe, M Littner, M Tagger, A Tamse.
Abstract
National and international standards require that gutta-percha cones be sufficiently radiopaque to be distinguished from natural structures such as dentin and to permit evaluation of the density of the root canal filling. Discs, 1-mm thick, made from 15 commercial and experimental brands of gutta-percha cones, were tested for their compliance with the requirements. The radiopacity of 1-mm thick root dentin slabs was similarly measured with a photo-densitometer. All radiographs were made on D-speed occlusal film and replicated on E-speed film. The mean radiopacity of the gutta-percha discs on D-speed film was 7.26 mm of aluminum equivalent and 7.53 mm on E-speed film, greatly exceeding the minimal requirement of 3 mm. The difference in the values obtained with the two film types was not significant, suggesting that E-speed films may be used for the test. Dentin slabs were uniform in their radiopacity, equivalent to 1 mm of aluminum, confirming previous findings. In view of complaints of insufficient clinical performance, the present minimal requirement for radiopacity of gutta-percha cones seems too low.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2074412 DOI: 10.1016/S0099-2399(06)81671-7
Source DB: PubMed Journal: J Endod ISSN: 0099-2399 Impact factor: 4.171