Unilateral placement of the Adiana device after failed bilateral placement of the Essure device for hysteroscopic sterilization.

Richard J Gimpelson1, Corey A Wagner.   

Abstract

This case report outlines the successful completion of hysteroscopic sterilization using the Adiana Permanent Contraception device (Hologic, Inc., Bedford, MA) after failure of multiple attempts to bilaterally place the Essure Permanent Birth Control system (Conceptus Inc., Mountain View, CA). The patient desired hysteroscopic sterilization using the Essure system. Although the first coil was easily placed in the right ostium, the second coil could not be passed beyond 2.4 cm into the left ostium. Inasmuch as the Adiana system needs to be passed only 1.4 cm into the ostium, this was found to be adequate for appropriate placement of the implant. Adiana may be an effective alternative to Essure, especially in patients in whom tubal stricture or spasm prevents Essure placement.
Copyright © 2010 AAGL. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20728829     DOI: 10.1016/j.jmig.2010.05.006

Source DB:  PubMed          Journal:  J Minim Invasive Gynecol        ISSN: 1553-4650            Impact factor:   4.137


  2 in total

1.  Successful placement of the Essure device after a failed procedure using the Adiana system for hysteroscopic sterilisation.

Authors:  Teska Schuurman; Sebastiaan Veersema
Journal:  BMJ Case Rep       Date:  2011-08-29

2.  Clinical performance characteristics of the adiana® system for permanent contraception: the first year of commercial use.

Authors:  Seth J Herbst; Edward G Evantash
Journal:  Rev Obstet Gynecol       Date:  2010
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.