| Literature DB >> 2072687 |
Abstract
In the absence of national or international electroacoustic standards for the evaluation of Frequency Modulated (FM) amplification systems, it becomes important to know the variability one may expect across similar models. Evaluation of thirty FM systems of the same model obtained from three different educational sites was performed to determine the variability that may occur as a result of the receiver, lapel microphone, or neckloop. There was a range as great as 20 dB in high frequency average saturation sound pressure level and equivalent input noise across receivers, lapel microphones, and neckloops. These results highlight the need for regular electroacoustic monitoring of not only the FM transmitter and receiver, but also the individual components, such as the lapel microphone and the neckloop.Mesh:
Year: 1991 PMID: 2072687 DOI: 10.1044/jshr.3403.628
Source DB: PubMed Journal: J Speech Hear Res ISSN: 0022-4685