Limitations of the DF-4 defibrillator connector necessitating device removal.

Gregory A Cogert1, Craig S Cameron, David A Sandler.   

Abstract

The DF-4 implantable defibrillator connector was recently released for clinical practice. This connector facilitates lead to device connection, reduces bulk in the device pocket, and eliminates the risk of incorrect device connection. Unfortunately, new technology often introduces new challenges. We report the case of a 63-year-old male with chronic systolic heart failure referred for cardiac resynchronization therapy-defibrillator implant. Limitations implicit to the current iteration of this technology include a lack of additional connectivity. In the present case, these limitations ultimately warranted device removal and reimplant with a traditional trifurcating IS-1/DF-1 connector. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20723081     DOI: 10.1111/j.1540-8159.2010.02873.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks.

Authors:  Christian Sticherling; Haran Burri
Journal:  Europace       Date:  2012-03-01       Impact factor: 5.214

2.  A Novel Solution for the High Defibrillation Threshold in Patients with a DF-4 Lead: Adding a High Voltage Adaptor/Splitter.

Authors:  Shmuel Inbar; Srikanth Seethala
Journal:  Indian Pacing Electrophysiol J       Date:  2014-05-25

3.  DF-4 Lead Connector: Innovative Technology, Unexpected Problems and Novel Solutions.

Authors:  Kartikeya Bhargava
Journal:  Indian Pacing Electrophysiol J       Date:  2014-05-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.