Literature DB >> 20723078

Performance of the lumenless 4.1-Fr diameter pacing lead implanted at alternative pacing sites in congenital heart: a chronic 5-year comparison.

Asra Khan1, Kathleen Zelin, Peter P Karpawich.   

Abstract

PURPOSE: United States approval of the Model 3830, 4.1-French (Fr) diameter, lumenless, pacing lead (Medtronic Inc., Minneapolis, MN, USA) in patients under 17 years of age, and those with congenital heart disease (CHD), was in 2005. To date, long-term performance at alternative pacing sites (APS) is limited and chronic efficacy comparisons with more established leads is lacking. The purpose of this study was to evaluate these factors.
METHODS: Implant and follow-up data on leads were compared: group 1 (non-3830 leads) and group 2 (Model 3830 leads). These included acute and chronic sensing and pacing, impedances, implant sites, and complications. Groups were compared using Fischer's exact test, paired, and nonpaired t-tests, with significance defined at P < 0.05.
RESULTS: A total of 119 patients (ages 5-48 years) received 171 leads: group 1 (n = 80) and group 2 (n = 91). At implant, there were no differences in patient age, CHD, sensing, or pacing thresholds between groups. Implant lead impedances differed between groups but all were within normal values for each lead design. Chronic data showed no difference in sensing, pacing thresholds, or impedances. There were five (6%) early lead dislodgements in group 1 and one (1%) in group 2. APS were achieved in group 2 with mean 1.6 ± 1.3 minutes fluoroscopy time.
CONCLUSION: The new 4.1-Fr lumenless lead shows similar performance indices to established leads even at APS, yet is thinner and achieves APS with technical ease, permitting more efficient chronic pacing in children and all patients with CHD. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20723078     DOI: 10.1111/j.1540-8159.2010.02884.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Pacing in children.

Authors:  Harinder R Singh; Anjan S Batra; Seshadri Balaji
Journal:  Ann Pediatr Cardiol       Date:  2013-01

2.  Case series of late lead dislodgement of Medtronic SelectSecure 3830 pacing leads in growing paediatric patients.

Authors:  Li Yen Ng; Sarah Gallagher; Kevin P Walsh
Journal:  Eur Heart J Case Rep       Date:  2021-01-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.