The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755.

A Jean1, C Boutet, B Lenormand, M-P Callat, G Buchonnet, V Barbay, J-P Basuyau, M Vasse.   

Abstract

INTRODUCTION: The analytical performance and the abnormality messages on differential (flags) of the new analyzer Beckman Coulter DxH 800 were compared with those of the LH 755.
METHODS: First, we evaluated the accuracy of the results of the DxH 800, in comparison with the LH 755, in 125 samples without alarm using unflagged sample results on both analyzers. Second, flagged samples on the LH 755 but not flagged by the DxH 800 were evaluated by flow cytometry for accuracy of the DxH 800 results. Finally, we evaluated the sensitivity and specificity of abnormality messages on differential given by the analyzers, in comparison with manual blood smears.
RESULTS: The correlation coefficients (R) for complete blood count parameters and differential demonstrated that the DxH 800 results were similar to that of LH 755. Excellent correlation coefficients between DxH 800 and flow cytometry results were found for white blood cell count (R = 0.985, n = 31), platelet count (R = 0.976, n = 51) and nucleated red blood cells (R = 0.966, n = 37). The overall performance showed an increased sensitivity (0.892) and specificity (0.864) of the flags on DxH 800 when compared to the LH 755 (0.846 and 0.733, respectively).
CONCLUSION: The DxH 800 provides reliable results and increases laboratory efficiency by reducing working time and costs associated with the optical validation of the results.
© 2010 Blackwell Publishing Ltd.

Entities:  

Mesh:

Year:  2010        PMID: 20718875     DOI: 10.1111/j.1751-553X.2010.01257.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  6 in total

1.  Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

Authors:  Shubham Rastogi; Sukesh Chandran Nair; Pandiyan Murugan; Asady Sukanya Sukumar; Joy J Mammen; Saravanan Mullai
Journal:  Pract Lab Med       Date:  2022-04-18

2.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

3.  Reliable, accurate determination of the leukocyte differential of leukopenic samples by using Hematoflow method.

Authors:  Yongjun Jo; Soo Hwa Kim; Kwangsang Koh; Jongmoon Park; Yang Bo Shim; Jihyang Lim; Yonggoo Kim; Yeon-Joon Park; Kyungja Han
Journal:  Korean J Lab Med       Date:  2011-06-28

4.  Mean cell volumes of neutrophils and monocytes are promising markers of sepsis in elderly patients.

Authors:  A-Jin Lee; Sang-Gyung Kim
Journal:  Blood Res       Date:  2013-09-25

5.  Interest of the cellular population data analysis as an aid in the early diagnosis of SARS-CoV-2 infection.

Authors:  Marc Vasse; Marie-Christine Ballester; Degnile Ayaka; Dmitry Sukhachev; Frédérique Delcominette; Florence Habarou; Emilie Jolly; Elena Sukhacheva; Tiffany Pascreau; Éric Farfour
Journal:  Int J Lab Hematol       Date:  2020-08-19       Impact factor: 3.450

Review 6.  Reviewing the value of leukocytes cell population data (CPD) in the management of sepsis.

Authors:  Eloísa Urrechaga
Journal:  Ann Transl Med       Date:  2020-08
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.