D-X Xu1, M Vachon, A Densmore, R Ma, A Delâge, S Janz, J Lapointe, Y Li, G Lopinski, D Zhang, Q Y Liu, P Cheben, J H Schmid.
Abstract
We report a silicon-on-insulator ring resonator biosensor array with one output port, using wavelength division multiplexing as the addressing scheme. With the use of on-chip referencing for environmental drift cancellation, simultaneous monitoring of multiplexed molecular bindings is demonstrated, with a resolution of 0.3 pg/mm(2) (40 ag of total mass) for protein concentrations over 4 orders of magnitude down to 20 pM. Reactions are measured over time periods as long as 3 h with high stability.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20717452 DOI: 10.1364/OL.35.002771