Literature DB >> 2071639

Trochanteric fixation by cable grip in hip replacement.

M A Ritter1, L E Eizember, E M Keating, P M Faris.   

Abstract

We used the stainless steel cable grip system described by Dall and Miles in 1983 to fix trochanters in 40 hips after total arthroplasty with trochanteric osteotomy. The cable broke in 32.5% of the hips; the trochanter failed to unite in 37.5%. Significantly more cables broke when placed inside the femoral canal than when the cable was placed round the femoral shaft (58% as against 9.5%, difference p less than 0.01). The high incidence of breakage may have resulted from contact between the stainless steel cable and the titanium prosthesis, from the acute angulation, or because of the lower fatigue strength of stainless steel. Better results have been obtained using cables with a higher fatigue strength, passed outside the proximal femur.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 2071639     DOI: 10.1302/0301-620X.73B4.2071639

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  14 in total

1.  The effects of femoral neck cut, cable tension, and muscles forces on the greater trochanter fixation.

Authors:  Yvan Petit; Luc P Cloutier; Kajsa Duke; G Yves Laflamme
Journal:  Med Biol Eng Comput       Date:  2012-03-09       Impact factor: 2.602

2.  Does trochanteric step osteotomy provide greater stability than classic slide osteotomy? A preliminary study.

Authors:  Ralf Schoeniger; Amy E LaFrance; Thomas R Oxland; Reinhold Ganz; Michael Leunig
Journal:  Clin Orthop Relat Res       Date:  2008-12-16       Impact factor: 4.176

3.  Greater trochanter displacements after Charnley's trochanterotomy, during passive hip movements: a real-time 3D cadaveric approach.

Authors:  Romain Gérard; Eric Stindel; Guillaume Dardenne; Jean-José Jacq; Romuald Seizeur; Bernard Sénécail; Christian Lefèvre
Journal:  Surg Radiol Anat       Date:  2009-03-14       Impact factor: 1.246

Review 4.  A systematic review and meta-analysis of complications following the posterior and lateral surgical approaches to total hip arthroplasty.

Authors:  J R Berstock; A W Blom; A D Beswick
Journal:  Ann R Coll Surg Engl       Date:  2015-01       Impact factor: 1.891

5.  [Osteotomy of the greater trochanter].

Authors:  K Anagnostakos; D Kohn; O Lorbach
Journal:  Orthopade       Date:  2013-05       Impact factor: 1.087

Review 6.  Extended trochanteric osteotomy: current concepts review.

Authors:  Senthil Nathan Sambandam; Gopinath Duraisamy; Jayadev Chandrasekharan; Varatharaj Mounasamy
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-02-29

7.  Reattachment of the greater trochanter in total hip arthroplasty: the pin-sleeve system compared with the Dall-Miles cable grip system.

Authors:  Naonobu Takahira; Moritoshi Itoman; Katsufumi Uchiyama; Sumitaka Takasaki; Kensuke Fukushima
Journal:  Int Orthop       Date:  2010-03-19       Impact factor: 3.075

8.  Early experience with a novel nonmetallic cable in reconstructive hip surgery.

Authors:  Nicholas T Ting; Glenn D Wera; Brett R Levine; Craig J Della Valle
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

9.  The role of trochanteric flip osteotomy in fixation of certain acetabular fractures.

Authors:  Sandeep Gupta; Jagdeep Singh; Jagandeep Singh Virk
Journal:  Chin J Traumatol       Date:  2017-03-23

10.  Use of locking plates for fixation of the greater trochanter in patients with hip replacement.

Authors:  Allison K Tetreault; Brian J McGrory
Journal:  Arthroplast Today       Date:  2016-10-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.