| Literature DB >> 20714891 |
K Hegnerová1, M Piliarik, M Šteinbachová, Z Flegelová, H Černohorská, J Homola.
Abstract
We present a compact surface plasmon resonance (SPR) biosensor for the detection of bisphenol A (BpA), an endocrine-disrupting chemical. The biosensor is based on an SPR sensor platform (SPRCD) and the binding inhibition detection format. The detection of BpA in PBS and wastewater was performed at concentrations ranging from 0.05 to 1,000 ng/ml. The limit of detection for BpA in PBS and wastewater was estimated to be 0.08 and 0.14 ng/ml, respectively. It was also demonstrated that the biosensor can be regenerated for repeated use. Results achieved with the SPR biosensor are compared with those obtained using ELISA and HPLC methods.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20714891 DOI: 10.1007/s00216-010-4067-z
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142