Literature DB >> 20708374

Acute aortic regurgitation secondary to disk embolization of a Björk-Shiley prosthetic aortic valve.

Robert D Grande1, William E Katz.   

Abstract

Having passed the 30th anniversary of the first implantation of a Björk-Shiley convexo-concave tilting mechanical valve, recognition of the life-threatening complication of strut fracture is not widespread. The authors report the case of a 48-year-old man with acute-onset chest pain and dyspnea found to have strut fracture and disk embolization of a 26-year-old Björk-Shiley prosthetic aortic valve. The value of echocardiography in the diagnosis of this condition is discussed.
Copyright © 2010 American Society of Echocardiography. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20708374     DOI: 10.1016/j.echo.2010.07.001

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  1 in total

1.  Outlet strut fracture and leaflet escape of Bjork-Shiley convexo-concave valve.

Authors:  Gaku Uchino; Hideo Yoshida; Naoya Sakoda; Shigeru Hattori; Takuya Kawabata; Munehiro Saiki; Yasufumi Fujita; Keiji Yunoki; Kunikazu Hisamochi; Yoshinari Mine
Journal:  Gen Thorac Cardiovasc Surg       Date:  2016-06-03
  1 in total