Force characteristics of nickel-titanium open-coil springs.

Allyson Bourke1, John Daskalogiannakis, Bryan Tompson, Philip Watson.   

Abstract

INTRODUCTION: The objective of this study was to quantify the properties of commercially available nickel-titanium open-coil springs.
METHODS: Eleven springs from 3 manufacturers were tested 5 times over a 12-week period. A universal testing machine was used to measure the force generated when open-coil springs were compressed to half of their original length and then gradually allowed to decompress.
RESULTS: The average forces generated at the initial recording session for uniformly wound springs from GAC International (Bohemia, NY) and 3M Unitek (Monrovia, Calif) were 19.3% to 42.7% and 9.7% to 38.8% below the manufacturers' labeled force levels, respectively. GAC's 100-, 150-, and 200-g stop-wound coils demonstrated statistically and clinically significant stepwise force degradation over the 12-week experimental period (P <0.0001). GAC's uniformly wound light (100 g) coils generated the lowest load-deflection ratios (23.7 g/mm).
CONCLUSIONS: Open coils might need to be compressed by more than one-third of their original length to produce the labeled forces. Uniformly wound coils generally produce lower load-deflection ratios and maximum forces, which are generally more acceptable for tooth movement. Copyright (c) 2010 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20691352     DOI: 10.1016/j.ajodo.2010.01.026

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  3 in total

1.  The effects of aging process and preactivation on mechanical properties of nickel-titanium closed coil springs.

Authors:  Shiva Alavi; Alireza Haerian
Journal:  Dent Res J (Isfahan)       Date:  2015 May-Jun

2.  A Comparison of Lower Dental Arch Changes Using Two Types of Space Regainers: A Removable Appliance with a Distalizing Screw and a Fixed Double-Banded Appliance.

Authors:  Shabnam Enteghad; Farzaneh Golfeshan; Ahmadreza Sardarian; Hooman Navaei
Journal:  Int J Dent       Date:  2022-04-16

3.  A Novel Alveolar Distractor Incorporating Nickel-Titanium Alloy Springs: A Preliminary In Vitro Study.

Authors:  Sarun Chancharoen; Peerapong Santiwong; Dutmanee Seriwatanachai; Anak Khantachawana; Rochaya Chintavalakorn
Journal:  Materials (Basel)       Date:  2022-07-25       Impact factor: 3.748

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.