| Literature DB >> 20687944 |
Negar Memarian1, Anastasios N Venetsanopoulos, Tom Chau.
Abstract
BACKGROUND: Recently, a novel single-switch access technology based on infrared thermography was proposed. The technology exploits the temperature differences between the inside and surrounding areas of the mouth as a switch trigger, thereby allowing voluntary switch activation upon mouth opening. However, for this technology to be clinically viable, it must be validated against a gold standard switch, such as a chin switch, that taps into the same voluntary motion.Entities:
Mesh:
Year: 2010 PMID: 20687944 PMCID: PMC2927594 DOI: 10.1186/1475-925X-9-38
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1Visual representation of the infrared thermal switch video processing algorithm. (a) Input grayscale thermal video frame, (b) Result of face localization, (c) Result of intensity analysis (warm area mask), (d) Intersection of warm area mask and motion mask, (e) Open mouth detected and marked on the video. Note that for ease of visualization, images (c)-(e) only show the smaller region demarcated by the dashed box in (b).
Figure 2Infrared thermal switch validity testing experiment setup.
Figure 3Efficiency of infrared thermal and chin switches averaged over all eighteen trials, for all ten participants. The vertical lines show standard deviation.
Selected qualitative feedback (pros and cons) from participants.
| Infrared Thermal Switch | Chin Switch | |
|---|---|---|
| Flexibility: switch works even in spite of user body motion and head rotation (as long as user face is in camera's field of view) | Instantaneous activation | |
| Non-contact | More sensitive switch | |
| Possible to activate the switch with closed eyes | ||
| Reduced detection sensitivity immediately after taking cold drink or food (this problem is alleviated with time) | Low detection sensitivity due to considerable head rotation or change of posture | |
| Reduced detection sensitivity if too much saliva in the mouth (this problem is alleviated with one swallow) | Too many false positives due to small head motion, talking or any activity that makes the chin touch the switch | |
| Prolonged periods of switch operation may cause fatigue or jaw pain | May cause infection if it enters user mouth | |