RASOR: an advanced instrument for soft x-ray reflectivity and diffraction.

T A W Beale1, T P A Hase, T Iida, K Endo, P Steadman, A R Marshall, S S Dhesi, G van der Laan, P D Hatton.   

Abstract

We report the design and construction of a novel soft x-ray diffractometer installed at Diamond Light Source. The beamline endstation RASOR is constructed for general users and designed primarily for the study of single crystal diffraction and thin film reflectivity. The instrument is comprised of a limited three circle (theta, 2theta, and chi) diffractometer with an additional removable rotation (phi) stage. It is equipped with a liquid helium cryostat, and post-scatter polarization analysis. Motorized motions are provided for the precise positioning of the sample onto the diffractometer center of rotation, and for positioning the center of rotation onto the x-ray beam. The functions of the instrument have been tested at Diamond Light Source, and initial test measurements are provided, demonstrating the potential of the instrument.

Entities:  

Year:  2010        PMID: 20687739     DOI: 10.1063/1.3458004

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523