| Literature DB >> 20687730 |
Seiji Higuchi1, Hiromi Kuramochi, Olivier Laurent, Takashi Komatsubara, Shinichi Machida, Masakazu Aono, Kenichi Obori, Tomonobu Nakayama.
Abstract
Over the past decade, multiple-scanning-probe microscope systems with independently controlled probes have been developed for nanoscale electrical measurements. We developed a quadruple-scanning-probe tunneling microscope (QSPTM) that can determine and control the probe position through scanning-probe imaging. The difficulty of operating multiple probes with submicrometer precision drastically increases with the number of probes. To solve problems such as determining the relative positions of the probes and avoiding of contact between the probes, we adopted sample-scanning methods to obtain four images simultaneously and developed an original control system for QSPTM operation with a function of automatic positional recognition. These improvements make the QSPTM a more practical and useful instrument since four images can now be reliably produced, and consequently the positioning of the four probes becomes easier owing to the reduced chance of accidental contact between the probes.Year: 2010 PMID: 20687730 DOI: 10.1063/1.3456990
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523