BACKGROUND: Correct positioning of the initial femoral guidewire is vital in order to prepare the femoral head properly for hip resurfacing. The purpose of the present investigation was to determine the accuracy and precision of the placement of the initial femoral guidewire with use of conventional alignment jigs and to compare the results with those of imageless computer navigation.
METHODS: Five commercially available jigs (two lateral pin jigs, two neck centering jigs, and one head planing jig) were obtained. Four surgeons used each jig and navigation three times to insert a guidewire in 10 degrees of relative valgus and neutral version into individual synthetic femora. A single surgeon then used each jig three times to align the initial guidewire in 10 degrees of relative valgus and neutral version in each of ten human cadaver femora. Radiographs of the synthetic and human femora were made to assess and compare guidewire inclination and version between conventional instrumentation and navigation.
RESULTS: Navigation provided ranges of error in the coronal guidewire alignment of up to eight times less than the conventional jigs, but both methods provided similar ranges of error for version. In both arms of the study, there were significant differences in coronal alignment accuracy between the two neck centering jigs. Next to navigation, one lateral pin jig provided the most accurate coronal placement of the initial guidewire whereas one neck centering jig provided the most precise coronal placement of the guidewire. Navigation was similar to conventional jigs in terms of the accuracy and precision of guidewire version.
CONCLUSIONS: In hip resurfacing arthroplasty, the choice of a femoral alignment device may influence the accuracy and precision of guidewire insertion, ultimately impacting femoral component placement. Imageless computer navigation can facilitate accurate and precise coronal alignment of the initial femoral guidewire, superior to that of conventional instrumentation. CLINICAL RELEVANCE: The results of this study may aid surgeons in the selection of alignment instruments for placement of the initial femoral guidewire during hip resurfacing.

Entities:  

Mesh:

Year:  2010        PMID: 20686057     DOI: 10.2106/JBJS.I.00648

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  3 in total

1.  Avoiding short-term femoral neck fracture with imageless computer navigation for hip resurfacing.

Authors:  Michael Olsen; Emil H Schemitsch
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

2.  Augmented Reality Based Navigation for Computer Assisted Hip Resurfacing: A Proof of Concept Study.

Authors:  He Liu; Edouard Auvinet; Joshua Giles; Ferdinando Rodriguez Y Baena
Journal:  Ann Biomed Eng       Date:  2018-05-23       Impact factor: 3.934

3.  Computer navigation experience in hip resurfacing improves femoral component alignment using a conventional jig.

Authors:  Zachary Morison; Akshay Mehra; Michael Olsen; Michael Donnelly; Emil Schemitsch
Journal:  Indian J Orthop       Date:  2013-11       Impact factor: 1.251

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.