Abstract
A novel, stable and highly sensitive non-enzymatic glucose (Glc) sensor was developed using vertically well-aligned multi-walled carbon nanotubes array (MWCNTs) incorporated with cupric oxide (CuO) nanoparticles. The MWCNTs array was prepared by catalytic chemical vapor deposition on a tantalum (Ta) substrate, while a simple and rapid two-step electrodeposition technique was used to prepare the CuO-MWCNTs nanocomposite. First, Cu nanoparticles were deposited onto MWCNTs at constant potential and then they were oxidized into CuO by potential cycling. The electrocatalytic activity of CuO-MWCNTs array was investigated for Glc under alkaline conditions using cyclic voltammetry and chronoamperometry. The sensor exhibited a linear response up to 3 mM of Glc and sensitivity of 2190 microA mM(-1) cm(-2), which is two to three orders of magnitude higher than that of most non-enzymatic Glc sensors reported in the literature. The sensor response time is less than 2s and detection limit is 800 nM (at signal/noise=3). When tested with human blood serum samples, the sensor exhibited high electrocatalytic activity, stability, fast response and good selectivity against common interfering species, suggesting its potential to be developed as a non-enzymatic Glc sensor. Copyright 2010 Elsevier B.V. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20685430 DOI: 10.1016/j.talanta.2010.03.047
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057