Influence of vessel size and tortuosity on in-stent restenosis after stent implantation in the vertebral artery ostium.

Zhiming Zhou1, Qin Yin, Gelin Xu, Xuanye Yue, Renliang Zhang, Wusheng Zhu, Xiaobing Fan, Minmin Ma, Xinfeng Liu.   

Abstract

PURPOSE: Percutaneous transluminal angioplasty and stenting is emerging as an alternative for treating atherosclerotic stenosis in the vertebral artery ostium. However, in-stent restenosis (ISR) still remains a critical issue to be addressed. Little is known about the relationship between anatomic characteristics of the artery and ISR after stent implantation. In this study, we have evaluated influential factors for ISR in a cohort of the patients with stenting in the vertebral artery ostium.
METHODS: Sixty-one patients with 63 symptomatic lesions in vertebral artery ostium treated with stenting were enrolled onto this study. An average of 12.5 months' clinical and angiographic follow-up results were analyzed retrospectively. The possible influential factors for ISR, including conventional risk factors of cerebrovascular diseases and morphological characteristics of target lesions, were evaluated by univariate and multivariate regression analysis.
RESULTS: Technical success was achieved in all 63 interventional procedures. Stenosis was reduced from (mean±standard deviation) 75.5±12% before to 1±3.6% after the procedure. During the mean 12.5-month angiographic follow-up, ISR was detected in 17 treated vessels (27.0%), with 2 treated arteries (3.2%) resulting in occlusion, and a stent fracture in 1 case (1.6%). Multivariate Cox regression analysis showed that the tortuosity of V1 (hazard ratio 3.54, P=0.01) and smaller diameter of the stent (hazard ratio 3.8, P=0.04) were independent predictors of ISR.
CONCLUSIONS: Angioplasty and stenting for symptomatic stenosis in the vertebral artery ostium stenosis seem to be feasible and effective. Tortuosity and smaller diameter may affect ISR after stent implantation.

Entities:  

Mesh:

Year:  2010        PMID: 20683721     DOI: 10.1007/s00270-010-9953-4

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  7 in total

1.  Endovascular treatment of extracranial vertebral artery stenosis.

Authors:  Burak Kocak; Bora Korkmazer; Civan Islak; Naci Kocer; Osman Kizilkilic
Journal:  World J Radiol       Date:  2012-09-28

2.  Long-Term Risk of In-Stent Restenosis and Stent Fracture for Extracranial Vertebral Artery Stenting.

Authors:  Man Kwun Andrew Li; Anderson Chun On Tsang; Frederick Chun Pong Tsang; Wai Shing Ho; Raymand Lee; Gilberto Ka Kit Leung; Wai Man Lui
Journal:  Clin Neuroradiol       Date:  2018-07-23       Impact factor: 3.649

3.  Stent fracture and occlusion after treatment of symptomatic vertebral artery ostium stenosis with a self-expanding device. A case report.

Authors:  Jun Lu; Jiachun Liu; Daming Wang; Shuo Wang
Journal:  Interv Neuroradiol       Date:  2014-12-05       Impact factor: 1.610

4.  Validation and comparison of drug eluting stent to bare metal stent for restenosis rates following vertebral artery ostium stenting: A single-center real-world study.

Authors:  Long Li; Xu Wang; Bin Yang; Yabing Wang; Peng Gao; Yanfei Chen; Fengshui Zhu; Yan Ma; Haitao Chi; Xiao Zhang; Xuesong Bai; Yao Feng; Adam A Dmytriw; Tao Hong; Yang Hua; Liqun Jiao; Feng Ling
Journal:  Interv Neuroradiol       Date:  2020-08-16       Impact factor: 1.610

5.  Effects of the Vertebral Artery Ostium/Subclavian Artery Angle on In-Stent Restenosis after Vertebral Artery Ostium Stenting.

Authors:  Hui Su; Shengyuan Yu; Chenglin Tian; Zhihua Du; Xinfeng Liu; Jun Wang; Xiangyu Cao
Journal:  Biomed Res Int       Date:  2021-04-27       Impact factor: 3.411

6.  Comparison of drug-eluting and bare metal stents for extracranial vertebral artery stenting.

Authors:  Damian R Maciejewski; Piotr Pieniazek; Lukasz Tekieli; Piotr Paluszek; Tadeusz Przewlocki; Tomasz Tomaszewski; Roman Machnik; Mariusz Trystula; Jacek Legutko; Anna Kablak-Ziembicka
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-09-18       Impact factor: 1.426

7.  Balloon-Assisted Angioplasty for the Treatment of In-Stent Restenosis After Vertebral Artery Ostium Stenting: Experiences From One Single Center.

Authors:  Xueli Cai; Xueping Chen; Yian Xiang; Liujing Chen; Jingping Sun
Journal:  Neurologist       Date:  2022-05-01       Impact factor: 1.524

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.