Very late thrombosis of a pipeline embolization device construct: case report.

David Fiorella1, Daniel Hsu, Henry H Woo, Robert W Tarr, Peter Kim Nelson.   

Abstract

BACKGROUND: The Pipeline embolization device (PED) is a new endoluminal construct designed to exclude aneurysms from the parent cerebrovasculature. We report the very late (>1 year) thrombosis of a PED construct placed for the treatment of a left vertebral aneurysm. CLINICAL
PRESENTATION: A patient with an occluded right vertebral artery and a large, fusiform intracranial left vertebral artery aneurysm was treated with PED and coil reconstruction. A durable, complete occlusion of the aneurysm was confirmed with control angiography at 1 year. The patient remained neurologically normal for 23 months until he experienced a transient visual disturbance followed weeks later by a minor brainstem stroke. INTERVENTION: Imaging evaluation showed thrombosis of the PED construct with complete occlusion of the left vertebral artery. After this stroke, he was initially treated with dual antiplatelet therapy and was then converted to warfarin. The patient remained neurologically stable for 5 months until he experienced progressive basilar thrombosis that ultimately resulted in a fatal stroke.
CONCLUSION: The PED represents a promising new endovascular technology for the treatment of cerebral aneurysms; however, as an investigational device, long-term follow-up data are sparse at this point. The etiology of the very late thrombosis of the PED construct in this case remains unknown; however, this report underscores the need for a continued, careful systematic evaluation and close long-term follow-up of treated patients.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20679914     DOI: 10.1227/01.NEU.0000383875.08681.23

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  50 in total

1.  Delayed ipsilateral parenchymal hemorrhage following flow diversion for the treatment of anterior circulation aneurysms.

Authors:  J P Cruz; M Chow; C O'Kelly; B Marotta; J Spears; W Montanera; D Fiorella; T Marotta
Journal:  AJNR Am J Neuroradiol       Date:  2012-03-08       Impact factor: 3.825

2.  Flow diversion for cerebral aneurysms: a cautionary tale.

Authors:  H J Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-11       Impact factor: 3.825

3.  Delayed spontaneous rupture of a posterior inferior cerebellar artery aneurysm following treatment with flow diversion: a clinicopathologic study.

Authors:  M Chow; C McDougall; C O'Kelly; R Ashforth; E Johnson; D Fiorella
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-11       Impact factor: 3.825

4.  Management strategy for bilateral complex vertebral artery aneurysms.

Authors:  Norihiro Saito; Hiroyasu Kamiyama; Katsumi Takizawa; Seiji Takebayashi; Takeshi Asano; Tohru Kobayashi; Rina Kobayashi; Shunsuke Kubota; Yasuhiro Ito; Kostadin L Karagiozov
Journal:  Neurosurg Rev       Date:  2015-11-13       Impact factor: 3.042

5.  Very late in-stent thrombosis 9 years after double stent treatment of fusiform basilar artery aneurysm.

Authors:  Robert Juszkat; Katarzyna Stanislawska; Norbert Wasik; Roman Jankowski; Włodzimierz Liebert
Journal:  Interv Neuroradiol       Date:  2015-05-11       Impact factor: 1.610

6.  Intravenous flat detector CT angiography for non-invasive visualisation of intracranial flow diverter: technical feasibility.

Authors:  Tobias Struffert; Marc Saake; Sabine Ott; Tobias Engelhorn; Philipp Gölitz; Stephan Kloska; Marc Doelken; Arnd Doerfler
Journal:  Eur Radiol       Date:  2011-03-30       Impact factor: 5.315

7.  Biodegradable flow-diverting device for the treatment of intracranial aneurysm: short-term results of a rabbit experiment.

Authors:  Kuizhong Wang; Shaoji Yuan; Xuping Zhang; Qiang Liu; Qisheng Zhong; Rongwei Zhang; Peigang Lu; Jiwen Li
Journal:  Neuroradiology       Date:  2013-02-05       Impact factor: 2.804

8.  Vertebro-basilar junction aneurysms: a single centre experience and meta-analysis of endovascular treatments.

Authors:  Francesca Graziano; Mario Ganau; Domenico Gerardo Iacopino; Edoardo Boccardi
Journal:  Neuroradiol J       Date:  2014-12-01

9.  Pipeline embolization device in aneurysmal subarachnoid hemorrhage.

Authors:  J P Cruz; C O'Kelly; M Kelly; J H Wong; W Alshaya; A Martin; J Spears; T R Marotta
Journal:  AJNR Am J Neuroradiol       Date:  2012-10-11       Impact factor: 3.825

10.  Evolution of Flow-Diverter Endothelialization and Thrombus Organization in Giant Fusiform Aneurysms after Flow Diversion: A Histopathologic Study.

Authors:  I Szikora; E Turányi; M Marosfoi
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-06       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.