Literature DB >> 20678914

Impedimetric immunosensor based on SWCNT-COOH modified gold microelectrodes for label-free detection of deep venous thrombosis biomarker.

S Bourigua1, M Hnaien, F Bessueille, F Lagarde, S Dzyadevych, A Maaref, J Bausells, A Errachid, N Jaffrezic Renault.   

Abstract

Measurement of D-dimer has subsequently become an essential element in the diagnostics of deep vein thrombosis and pulmonary embolism; in this context microelectrodes with an area of 9×10(-4) cm(2) were used to develop impedimetric immunosensor for detecting deep venous thrombosis biomarker (D-dimer). The biosensor is based on functionalized carbon nanotubes (SWCNT-COOH) where the antibody (anti-D-dimer) was immobilized by covalent binding. The electrical properties and the morphology of the biolayer were characterized by electrochemical impedance spectroscopy (EIS), cyclic voltammetry and atomic force spectroscopy (AFM). Impedimetric microimmunosensor allows to obtain sensitivity of 40.1 kΩ μM(-1) and detection limit of 0.1 pg/mL (0.53 fM) with linear range from 0.1 pg/mL to 2 μg/mL (0.53 fM to 0.01 μM). We demonstrate that using carbon nanotubes and microelectrodes, high sensitivity and dynamic range were obtained. The biosensor exhibited a short response time of 10 min. Moreover, the studied immunosensor exhibits good reproducibility (R.S.D. 8.2%, n=4).
Copyright © 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20678914     DOI: 10.1016/j.bios.2010.07.004

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  3 in total

Review 1.  Electrochemical sensing: A prognostic tool in the fight against COVID-19.

Authors:  Sharda Kotru; Martin Klimuntowski; Hashim Ridha; Zakir Uddin; Ali A Askhar; Gurmit Singh; Matiar M R Howlader
Journal:  Trends Analyt Chem       Date:  2021-01-23       Impact factor: 14.908

2.  Sensitive impedimetric biosensor for direct detection of diazinon based on lipases.

Authors:  Nedjla Zehani; Sergei V Dzyadevych; Rochdi Kherrat; Nicole J Jaffrezic-Renault
Journal:  Front Chem       Date:  2014-07-04       Impact factor: 5.221

Review 3.  A review on impedimetric immunosensors for pathogen and biomarker detection.

Authors:  J Leva-Bueno; Sally A Peyman; P A Millner
Journal:  Med Microbiol Immunol       Date:  2020-04-03       Impact factor: 3.402

  3 in total