Literature DB >> 20671629

Use of the AorfixTM stent graft in patients with tortuous iliac anatomy.

A R Weale1, K Balasubramaniam, J Hardman, M Horrocks.   

Abstract

AIM: Iliac limb occlusion following endovascular repair (EVAR) may result in limb threatening acute ischemia. The incidence is reported at up to 10% and is known to be influenced by iliac angulation and kinking of the stent graft. The aim of this study was to evaluate the performance of the AorfixTM graft in tortuous iliac anatomy and examine the impact of the graft on the rate of iliac limb thrombosis following EVAR in a single UK centre
METHODS: We performed a retrospective review of all EVAR performed from May 1998 to May 2010. From November 2007, patients with highly angulated iliac anatomy were treated with the AorfixTM(Lombard) stent graft, or when a ZenithTM(Cook) main body was chosen, the AorfixTM iliac limbs were used with the ZenithTM(Cook) device. We compared the rate of iliac limb occlusions before (group 1) and after (group 2) the adoption of this policy.
RESULTS: Two hundred and sixty seven patients underwent EVAR (group 1 n=129; group 2 n=138). In group 1, eight patients had a unilateral iliac limb occlusion (6.2%). Six of the patients had >90 degrees iliac angulation, one had an unrecognised limb stenosis, and one patient had the stent landed in the external iliac. In group 2 there were no limb occlusions. Of the 138 patients, 25 had iliac angulation of >90 degrees . Of these 25, eighteen were treated with the AorfixTM stent graft system because of iliac angulation, and 7 were treated with AorfixTM legs and ZenithTM bodies.
CONCLUSION: The rate of early iliac limb occlusion following EVAR in patients with angulated iliac anatomy can be substantially reduced by using the flexible Aorfix stent graft system.

Entities:  

Mesh:

Year:  2010        PMID: 20671629

Source DB:  PubMed          Journal:  J Cardiovasc Surg (Torino)        ISSN: 0021-9509            Impact factor:   1.888


  3 in total

1.  Treatment of Hostile Proximal Necks During Endovascular Aneurysm Repair.

Authors:  Tulio Pinho Navarro; Rodrigo de Castro Bernardes; Ricardo Jayme Procopio; Jose Oyama Leite; Alan Dardik
Journal:  Aorta (Stamford)       Date:  2014-02-01

2.  Mechanical characterizations of braided composite stents made of helical polyethylene terephthalate strips and NiTi wires.

Authors:  Qingli Zheng; Pengfei Dong; Zhiqiang Li; Xinwei Han; Changchun Zhou; Meiwen An; Linxia Gu
Journal:  Nanotechnol Rev       Date:  2019-11-06       Impact factor: 6.739

3.  Braided composite stent for peripheral vascular applications.

Authors:  Qingli Zheng; Pengfei Dong; Zhiqiang Li; Ying Lv; Meiwen An; Linxia Gu
Journal:  Nanotechnol Rev       Date:  2020-11-27       Impact factor: 6.739