Literature DB >> 20664378

The effects of magnetic resonance imaging on surgical staples: an experimental analysis.

J Christopher Gayton1, Paul Sensiba, Brian F Imbrogno, Indresh Venkatarayappa, James Tsatalis, Michael J Prayson.   

Abstract

BACKGROUND: Surgical staples are commonplace in repairing surgical incisions. Staples allow for expeditious closure and removal compared with suture materials. However, there are clinical concerns when obtaining a magnetic resonance imaging (MRI) scan with staples present. This study examined two issues related to MRI scanning in the presence of surgical staples: skin surface temperature change and staple displacement.
METHODS: Thirty pig feet had 3-cm surgical incisions repaired with five surgical staples. Once placed, each skin staple position was marked for later referencing. A surface temperature laser device recorded prescan skin surface temperature. A 35-minute MRI scan was performed with a 1.5-Tesla magnet and standard knee coil for each pig foot. Immediately afterward, the skin surface temperature and displacement measurements were recorded. The paired t test was used to analyze temperature change from prescan to postscan.
RESULTS: The prescan mean temperature was 16.45°C (standard deviation: 0.70°C), and the range was 14.60°C to 18.20°C. After scanning, the mean temperature was 16.02°C (standard deviation: 0.63°C), and the range was 15.00°C to 17.60°C. The decrease of 0.43°C in skin surface temperature was statistically significant (p=0.001). No change in staple position was measurable or evident by visual inspection for any of the pig feet.
CONCLUSION: This study found no increase in skin surface temperature or displacement of staple position after a standard extremity MRI scan. Based on our findings, MRI scanning in the presence of stainless steel surgical staples seems safe.

Entities:  

Mesh:

Year:  2011        PMID: 20664378     DOI: 10.1097/TA.0b013e3181de3855

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  2 in total

1.  Epidural Hematoma: Vigilance beyond Guidelines.

Authors:  Nitin Madhukar Bhorkar; Tasneem Saleh Dhansura; Urmila Bhaktiprasad Tarawade; Sanket Sharad Mehta
Journal:  Indian J Crit Care Med       Date:  2018-07

Review 2.  MRI for an Acute Secondary Site Complication in Post-arthroplasty Management: Narrative Review of Safety Concerns for an Implanted Hip and Knee Joint.

Authors:  Lavindra Tomar; Gaurav Govil; Pawan Dhawan
Journal:  Cureus       Date:  2022-02-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.