Four color-measuring devices compared with a spectrophotometric reference system.

Karl Martin Lehmann1, Christopher Igiel, Irene Schmidtmann, Herbert Scheller.   

Abstract

OBJECTIVES: The objective of this study was to compare the L*C*h° color coordinates of dental color-measuring devices with those of a spectrophotometric reference system.
METHODS: The International Commission on Illumination (CIE) L*C*h° color coordinates of ceramic samples, matching the colors of the VITA Linearguide, were recorded using four color-measuring devices (VITA Easyshade (A), VITA Easyshade compact (B), Degudent Shadepilot (C), X-Rite Shadevision (D)) and a spectrophotometric reference system under standardized test conditions. The intraclass correlation coefficients, regression lines with coefficients of determination, and mean deviations of the dental color-measuring devices from the spectrophotometric reference system were calculated.
RESULTS: All of the devices had high intraclass correlation coefficients, between 0.979 and 1.000. Nearly all measurements, except the hue values of devices A, B, and D, showed coefficients of determination close to 1.0 over the range of measurements for L*C*h° color coordinates. Similarly, all of the devices had coefficients of determination near the optimum value of 1.0. Devices C and D produced regression line slopes near the optimum value of 1.0 and intercepts close to the optimum value of zero. Only the L* coordinate measurements of devices A and B and the h° coordinate measurements of device B differed not significantly from the spectrophotometric reference values.
CONCLUSIONS: The electronic dental color-measuring devices tested showed excellent repeatability, but some devices showed substantial deviations in color coordinate values from the spectrophotometric reference system. Devices C and D showed higher precision than devices A and B.
Copyright © 2010 Elsevier Ltd. All rights reserved.

Mesh:

Year:  2010        PMID: 20659526     DOI: 10.1016/j.jdent.2010.07.006

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  16 in total

1.  Reliability of shade selection using an intraoral spectrophotometer.

Authors:  Siegbert Witkowski; Nao-Daniel Yajima; Martin Wolkewitz; Jorge R Strub
Journal:  Clin Oral Investig       Date:  2011-08-10       Impact factor: 3.573

2.  Interdevice agreement of eight equivalent dental color measurement devices.

Authors:  M Weyhrauch; C Igiel; A M Pabst; S Wentaschek; H Scheller; K M Lehmann
Journal:  Clin Oral Investig       Date:  2015-03-24       Impact factor: 3.573

3.  In vivo and in vitro spectrophotometric evaluation of upper central incisors before and after extraction.

Authors:  Enric Soler; Fernando Duran-Sindreu; Juan Basilio; Miguel Roig; Stefano Ardu; Juan R Mayoral
Journal:  Clin Oral Investig       Date:  2017-01-05       Impact factor: 3.573

4.  The repeatability of digital shade measurement--a clinical study.

Authors:  Constanze Olms; Jürgen M Setz
Journal:  Clin Oral Investig       Date:  2012-07-25       Impact factor: 3.573

5.  Spectrophotometric and computerized evaluation of tooth bleaching employing 10 different home-bleaching procedures: In-vitro study.

Authors:  Cem Peskersoy; Ayhan Tetik; Veli Ozgen Ozturk; Necmi Gokay
Journal:  Eur J Dent       Date:  2014-10

6.  Comparison of accuracies of an intraoral spectrophotometer and conventional visual method for shade matching using two shade guide systems.

Authors:  Vidhya Parameswaran; S Anilkumar; S Lylajam; C Rajesh; Vivek Narayan
Journal:  J Indian Prosthodont Soc       Date:  2016 Oct-Dec

7.  Tooth Discoloration Resulting from a Nano Zinc Oxide-Eugenol Sealer.

Authors:  Mina Zarei; Maryam Javidi; Marzieh Jafari; Maryam Gharechahi; Pedram Javidi; Maryam Shayani Rad
Journal:  Iran Endod J       Date:  2017

8.  Bovine Tooth Discoloration Induced by Endodontic Filling Materials for Primary Teeth.

Authors:  Samantha Rodrigues Xavier; Katerine Jahnecke Pilownic; Andressa Heberle Gastmann; Mariana Silveira Echeverria; Ana Regina Romano; Fernanda Geraldo Pappen
Journal:  Int J Dent       Date:  2017-04-05

9.  Spectrophotometric color evaluation of permanent incisors, canines and molars. A cross-sectional clinical study.

Authors:  Ioana-Sofia Pop-Ciutrila; Horatiu Alexandru Colosi; Diana Dudea; Mandra Eugenia Badea
Journal:  Clujul Med       Date:  2015-11-15

10.  Perceived color shift of ceramics according to the change of illuminating light with spectroradiometer.

Authors:  Hyun-Suk Cha; Bin Yu; Yong-Keun Lee
Journal:  J Adv Prosthodont       Date:  2013-08-31       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.