Effects of different fissure sealant applications on laser fluorescence measurements.

Işll Saroğlu Sönmez1, Aylin Akbay Oba, Merve Erkmen, Seda Ekici.   

Abstract

OBJECTIVE: The aim of this in vitro study was to evaluate the effects of using only phosphoric acid or a self-etch bonding agent under clear and opaque fissure sealants on laser fluorescence (LF) readings and the reproducibility of the laser device.
METHODS: Eighty extracted permanent molars, ranged from sound to carious, were randomly divided into four groups: phosphoric acid + opaque sealant (group I), Clearfil S3 Bond (Kuraray, Kurashiki, Japan) + opaque sealant (group II), phosphoric acid + clear sealant (group III), and Clearfil S3 Bond + clear sealant (group IV). The teeth were measured using an LFpen device, before and after sealing. Data were analysed using the Spearman's correlation, Wilcoxon signed rank, and Mann-Whitney U-test.
RESULTS: Except group IV, there was a statistically significant decrease in fluorescence after the application of sealants (P < 0.05). The decrease of LFpen readings in the opaque sealant groups was more significant than the clear sealant groups (P < 0.05). But for both sealants, the difference between phosphoric acid and Clearfil S3 Bond groups was nonsignificant (P > 0.05).
CONCLUSIONS: There was a statistically significant decrease in fluorescence for both clear and opaque sealant groups. However, clear sealant with Clearfil S3 Bond does not influence the LFpen readings.
© 2010 The Authors. International Journal of Paediatric Dentistry © 2010 BSPD, IAPD and Blackwell Publishing Ltd.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 20659181     DOI: 10.1111/j.1365-263X.2010.01076.x

Source DB:  PubMed          Journal:  Int J Paediatr Dent        ISSN: 0960-7439            Impact factor:   3.455


  2 in total

1.  Effects of tetracycline hydrochloride on measurements with the laser fluorescence device DIAGNOdent: in vitro and in vivo studies.

Authors:  Eriko Nagaoka; Tomio Morohashi; Junichirou Kinoshita; Akiko Karakawa; Nobuhiro Sakai; Shoji Yamada; Mitsuko Inoue
Journal:  Lasers Med Sci       Date:  2011-02-09       Impact factor: 3.161

2.  The effects of opaque and clear pit and fissure sealants on infrared laser fluorescence measurements.

Authors:  Z Bahrololoomi; M Khodabakhsh; Y Khaksar
Journal:  J Dent (Shiraz)       Date:  2014-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.