Electromagnetic real-time tumor position monitoring and dynamic multileaf collimator tracking using a Siemens 160 MLC: geometric and dosimetric accuracy of an integrated system.

Andreas Krauss1, Simeon Nill, Martin Tacke, Uwe Oelfke.   

Abstract

PURPOSE: Dynamic multileaf collimator tracking represents a promising method for high-precision radiotherapy to moving tumors. In the present study, we report on the integration of electromagnetic real-time tumor position monitoring into a multileaf collimator-based tracking system. METHODS AND MATERIALS: The integrated system was characterized in terms of its geometric and radiologic accuracy. The former was assessed from portal images acquired during radiation delivery to a phantom in tracking mode. The tracking errors were calculated from the positions of the tracking field and of the phantom as extracted from the portal images. Radiologic accuracy was evaluated from film dosimetry performed for conformal and intensity-modulated radiotherapy applied to different phantoms moving on sinusoidal trajectories. A static radiation delivery to the nonmoving target served as a reference for the delivery to the moving phantom with and without tracking applied.
RESULTS: Submillimeter tracking accuracy was observed for two-dimensional target motion despite the relatively large system latency of 500 ms. Film dosimetry yielded almost complete recovery of a circular dose distribution with tracking in two dimensions applied: 2%/2 mm gamma-failure rates could be reduced from 59.7% to 3.3%. For single-beam intensity-modulated radiotherapy delivery, accuracy was limited by the finite leaf width. A 2%/2 mm gamma-failure rate of 15.6% remained with tracking applied.
CONCLUSION: The integrated system we have presented marks a major step toward the clinical implementation of high-precision dynamic multileaf collimator tracking. However, several challenges such as irregular motion traces or a thorough quality assurance still need to be addressed. Copyright Â
© 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20656420     DOI: 10.1016/j.ijrobp.2010.03.043

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  18 in total

1.  Image-based dynamic multileaf collimator tracking of moving targets during intensity-modulated arc therapy.

Authors:  Per Rugaard Poulsen; Walther Fledelius; Byungchul Cho; Paul Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-06       Impact factor: 7.038

2.  Detailed analysis of latencies in image-based dynamic MLC tracking.

Authors:  Per Rugaard Poulsen; Byungchul Cho; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

3.  Time-resolved dose distributions to moving targets during volumetric modulated arc therapy with and without dynamic MLC tracking.

Authors:  Thomas Ravkilde; Paul J Keall; Cai Grau; Morten Høyer; Per R Poulsen
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

4.  Implementation and experimental results of 4D tumor tracking using robotic couch.

Authors:  I Buzurovic; Y Yu; M Werner-Wasik; T Biswas; P R Anne; A P Dicker; T K Podder
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

5.  Technical Note: In silico and experimental evaluation of two leaf-fitting algorithms for MLC tracking based on exposure error and plan complexity.

Authors:  Vincent Caillet; Ricky O'Brien; Douglas Moore; Per Poulsen; Tobias Pommer; Emma Colvill; Amit Sawant; Jeremy Booth; Paul Keall
Journal:  Med Phys       Date:  2019-03-04       Impact factor: 4.071

6.  The first clinical implementation of electromagnetic transponder-guided MLC tracking.

Authors:  Paul J Keall; Emma Colvill; Ricky O'Brien; Jin Aun Ng; Per Rugaard Poulsen; Thomas Eade; Andrew Kneebone; Jeremy T Booth
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

7.  Using an external surrogate for predictor model training in real-time motion management of lung tumors.

Authors:  Joerg Rottmann; Ross Berbeco
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

8.  Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors.

Authors:  J Rottmann; P Keall; R Berbeco
Journal:  Phys Med Biol       Date:  2013-05-28       Impact factor: 3.609

9.  Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Authors:  Lei Zhang; Thomas LoSasso; Pengpeng Zhang; Margie Hunt; Gig Mageras; Grace Tang
Journal:  Med Phys       Date:  2020-09-11       Impact factor: 4.071

10.  Fast leaf-fitting with generalized underdose/overdose constraints for real-time MLC tracking.

Authors:  Douglas Moore; Dan Ruan; Amit Sawant
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.