Retinal safety of the irradiation delivered to light-adjustable intraocular lenses evaluated in a rabbit model.

Liliana Werner1, Wellington Chang, Jonathan Haymore, Brian Haugen, Natalya Romaniv, Christian Sandstedt, Shiao Chang, Nick Mamalis.   

Abstract

PURPOSE: To evaluate the safety to the retina of a light-delivery device used to irradiate a light-adjustable intraocular lens (IOL) after implantation in a rabbit model.
SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
METHODS: In this study, rabbits had phacoemulsification with implantation of an ultraviolet (UV)-filtering light-adjustable IOL (study IOL) in 1 eye and a custom-made silicone IOL without a UV filter (control IOL) in the opposite eye. The study IOLs were irradiated at 1.0, 2.0, 3.0, and 5.0 times the expected maximum UV irradiation doses and the control IOLs, at 0.3, 0.6, 1.0, and 2.0 times. One week after irradiation, slitlamp and fundus (indirect ophthalmoscopy) examinations were performed. The rabbits were then humanely killed and their eyes enucleated and processed for histopathology.
RESULTS: The 16 eyes with the study IOL (with UV filter) showed no signs of corneal, anterior segment, or retinal toxicity on histopathologic evaluation. The 16 eyes with the control IOL (no UV filter) also showed no signs of corneal or anterior segment toxicity; however, 3 eyes receiving the higher radiation doses had focal areas of retinal damage consistent with laser burn.
CONCLUSION: Pigmented rabbit eyes with a light-adjustable IOL with a UV filter showed no signs of retina toxicity after near-UV light exposure up to 5 times the expected maximum treatment dosage. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned. Additional disclosures are found in the footnotes. Copyright 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20656165     DOI: 10.1016/j.jcrs.2010.02.018

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

1.  [Spectral transmission in blue filter intraocular lenses].

Authors:  N Schrage
Journal:  Ophthalmologe       Date:  2012-02       Impact factor: 1.059

Review 2.  Evaluating the evidence for and against the use of IOLs in infants and young children.

Authors:  Priyanka Kumar; Scott R Lambert
Journal:  Expert Rev Med Devices       Date:  2016-02-29       Impact factor: 3.166

3.  Evaluation of a 345 nm Femtosecond Laser for Corneal Surgery with Respect to Intraocular Radiation Hazard.

Authors:  Johannes Menzel-Severing; Corinna Petsch; Theofilos Tourtas; Naresh Polisetti; Jörg Klenke; Katrin Skerl; Christian Wüllner; Christof Donitzky; Friedrich E Kruse; Jan Kremers; Christian M Hammer
Journal:  PLoS One       Date:  2015-09-11       Impact factor: 3.240

4.  Oxidative stress and histological changes in a model of retinal phototoxicity in rabbits.

Authors:  Manuel Saenz-de-Viteri; Henar Heras-Mulero; Patricia Fernández-Robredo; Sergio Recalde; María Hernández; Nicholas Reiter; Maite Moreno-Orduña; Alfredo García-Layana
Journal:  Oxid Med Cell Longev       Date:  2014-05-27       Impact factor: 6.543

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.