Literature DB >> 20647924

Liver-assist device with a microfluidics-based vascular bed in an animal model.

Wen-Ming Hsu1, Amedeo Carraro, Katherine M Kulig, Mark L Miller, Mohammad Kaazempur-Mofrad, Eli Weinberg, Fateh Entabi, Hassan Albadawi, Michael T Watkins, Jeffrey T Borenstein, Joseph P Vacanti, Craig Neville.   

Abstract

OBJECTIVE: This study evaluates a novel liver-assist device platform with a microfluidics-modeled vascular network in a femoral arteriovenous shunt in rats. SUMMARY OF BACKGROUND DATA: Liver-assist devices in clinical trials that use pumps to force separated plasma through packed beds of parenchymal cells exhibited significant necrosis with a negative impact on function.
METHODS: Microelectromechanical systems technology was used to design and fabricate a liver-assist device with a vascular network that supports a hepatic parenchymal compartment through a nanoporous membrane. Sixteen devices with rat primary hepatocytes and 12 with human HepG2/C3A cells were tested in athymic rats in a femoral arteriovenous shunt model. Several parenchymal tube configurations were evaluated for pressure profile and cell survival. The blood flow pattern and perfusion status of the devices was examined by laser Doppler scanning. Cell viability and serum protein secretion functions were assessed.
RESULTS: Femoral arteriovenous shunt was successfully established in all animals. Blood flow was homogeneous through the vascular bed and replicated native flow patterns. Survival of seeded liver cells was highly dependent on parenchymal chamber pressures. The tube configuration that generated the lowest pressure supported excellent cell survival and function.
CONCLUSIONS: This device is the first to incorporate a microfluidics network in the systemic circulatory system. The microvascular network supported viability and function of liver cells in a short-term ex vivo model. Parenchymal chamber pressure generated in an arteriovenous shunt model is a critical parameter that affects viability and must be considered in future designs. The microfluidics-based vascular network is a promising platform for generating a large-scale medical device capable of augmenting liver function in a clinical setting.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20647924     DOI: 10.1097/SLA.0b013e3181e982ba

Source DB:  PubMed          Journal:  Ann Surg        ISSN: 0003-4932            Impact factor:   12.969


  6 in total

1.  What is the purpose of launching the World Journal of Transplantation?

Authors:  Maurizio Salvadori
Journal:  World J Transplant       Date:  2011-12-24

Review 2.  Liver bioengineering: from the stage of liver decellularized matrix to the multiple cellular actors and bioreactor special effects.

Authors:  Mireia Caralt; Enrique Velasco; Angel Lanas; Pedro M Baptista
Journal:  Organogenesis       Date:  2014-08-11       Impact factor: 2.500

3.  Using reconfigurable microfluidics to study the role of HGF in autocrine and paracrine signaling of hepatocytes.

Authors:  Dipali Patel; Amranul Haque; Yandong Gao; Alexander Revzin
Journal:  Integr Biol (Camb)       Date:  2015-07       Impact factor: 2.192

4.  Multiscale computational model of fluid flow and matrix deformation in decellularized liver.

Authors:  Kenichiro Nishii; Greg Reese; Emma C Moran; Jessica L Sparks
Journal:  J Mech Behav Biomed Mater       Date:  2015-12-07

Review 5.  Engineered liver for transplantation.

Authors:  Basak E Uygun; Martin L Yarmush
Journal:  Curr Opin Biotechnol       Date:  2013-06-20       Impact factor: 9.740

6.  Application of microtechnologies for the vascularization of engineered tissues.

Authors:  Robert Gauvin; Maxime Guillemette; Mehmet Dokmeci; Ali Khademhosseini
Journal:  Vasc Cell       Date:  2011-10-31
  6 in total