The influence of self-absorption on PET and PET/CT shielding requirements.

M Elschot1, T C de Wit, H W A M de Jong.   

Abstract

PURPOSE: The high energy (511 keV) annihilation photons used in positron emission tomography (PET) imaging generally require a substantial amount of lead to protect personnel and the general public from ionizing radiation. A cost-effective design of the PET facility that ensures radiation does not exceed formal dose limits requires accurate estimation of the necessary PET shielding. The American Association of Physicists in Medicine (AAPM) Task Group 108 recently published broad beam transmission factors based on Monte Carlo calculations of 511 keV photons. In this work, an extension to the AAPM model is presented, based on Monte Carlo simulations including the effects of self-absorption on the photon energy spectrum.
METHODS: Monte Carlo calculations were performed using MCNPX. The photon energy spectrum after self-absorption was computed by simulating a normal 18FDG activity distribution in an anthropomorphic phantom. This spectrum was used to calculate the dose rate transmission factors for various wall thicknesses of lead, concrete, and iron. The method was validated by measurement and corresponding simulation of the transmission factors of an 18FDG source in air and in PMMA. Furthermore, a method to generate 3D area dose rate maps of PET facilities incorporating the calculated transmission tables is presented and applied to several shielding situations.
RESULTS: The calculated self-absorption correction factor and the broad beam transmission factors resulting from Monte Carlo simulations of a monoenergetic point source emitting 511 keV photons were in excellent agreement with the results of the AAPM publication (0.66 vs 0.64 and R2 = 0.999, respectively). However, when all radiation physics, i.e., also the effect of self-absorption on the photon energy spectrum, is included in the Monte Carlo calculations, a substantial reduction in required shielding material was found. For example, including all radiation physics leads to 13.3 mm of lead required to obtain a typical transmission factor of 0.1, instead of 16.0 mm of lead when the AAPM data including only the self-absorption correction factor are used. These findings were confirmed by the experimental measurements. The transmission factors produced in this work can be applied in the same manner as those estimated by AAPM to allow for a cost-effective design of PET and PET/CT facilities without violating radiation safety regulations.
CONCLUSIONS: Taking into account the effect of self-absorption on the photon energy spectrum results in more accurate and cost-effective shielding requirement estimations.

Entities:  

Mesh:

Year:  2010        PMID: 20632612     DOI: 10.1118/1.3432618

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Estimation of ambient dose equivalent distribution in the 18F-FDG administration room using Monte Carlo simulation.

Authors:  Shuji Nagamine; Toshioh Fujibuchi; Yoshiyuki Umezu; Kazuhiko Himuro; Shinichi Awamoto; Yuji Tsutsui; Yasuhiko Nakamura
Journal:  Radiol Phys Technol       Date:  2016-08-16

2.  Effective dose to staff members in a positron emission tomography/CT facility using zirconium-89.

Authors:  K S Alzimami; A K Ma
Journal:  Br J Radiol       Date:  2013-08-09       Impact factor: 3.039

Review 3.  Translating a radiolabeled imaging agent to the clinic.

Authors:  Gary L Griffiths; Crystal Vasquez; Freddy Escorcia; Jeff Clanton; Liza Lindenberg; Esther Mena; Peter L Choyke
Journal:  Adv Drug Deliv Rev       Date:  2021-12-20       Impact factor: 15.470

4.  Investigation of 18F and 89Zr Isotopes Self-Absorption and Dose Rate Parameters for PET Imaging.

Authors:  Abdulrahman A Alfuraih; Khalid Alzimami; Andy K Ma
Journal:  Dose Response       Date:  2021-07-09       Impact factor: 2.658